9b96906150192f3f99585db331ef8574097f99fa
The radial parallax broadening is the variance of the depth at which a photon converts, so it scales with the attenuation length. That length was approximated as photoelectric-dominated and scaled by lambda^3 from a single 13 keV reference per material. For silicon above 10 keV that is within 0.2%, but for CdTe it overstates the attenuation length by up to a factor of two, and by six above the cadmium K edge - which made this variance term 1.9x too large on 750 um CdTe data. Take the length from the same tabulated coefficients the efficiency correction uses. Silicon data is unaffected to within the approximation's own error; CdTe data gets a spot-width variance that matches the sensor it was recorded on. Co-Authored-By: Claude Opus 5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01EFEJG6WBQv8th4UJFNe53N