feat(flomni): fix tomo-queue move allow-list, drop idgap stub, add foptz+shift action
CI for csaxs_bec / test (push) Successful in 3m10s

Queue-control "command job" changes:
- Remove the "optimize_idgap" stub action (not commissioned, always a
  no-op) from both Flomni's and LamNI's action registry, and delete the
  now-dead TomoQueueMixin.optimize_idgap() method.
- Fix the "move" action's device allow-list: "mokev" doesn't exist in the
  current device config at all -- the real energy device is "ccm_energy"
  (LamNI's copy already had this right). Also drop "ftray" (sample
  transfer tray), which was only ever there as a sim-testable move target.
- Add a new "move_foptz_with_shift" action: moves foptz (the FZP
  optics-Z stage) and sets the compensating manual_shift_x/manual_shift_y
  in one atomic step. foptz is deliberately NOT in the plain move
  allow-list -- moving it shifts the aligned scan center (interferometer
  mirrors aren't perfectly parallel to the beam), so this is the only way
  to queue a foptz move, specifically so the compensating shift can never
  be forgotten.
- The foptz target is validated against the device's real, live limits
  (not a hand-duplicated static number) both in the GUI spinbox's range
  and in the CLI's add-time validation, so a queued job can't raise when
  it executes unattended later; a defense-in-depth check also exists at
  execution time.

Retargeted the e2e command-job tests from ftray to foptz (now the only
sim-testable move-adjacent path) and updated the AI_docs testing/GUI-testing
notes to match the new action set.

Also fixes a pre-existing, unrelated broken import
(plugins.omny.omny_general_tools -> plugins.OMNY_shared.omny_general_tools,
already broken on main) in _bootstrap.py/test_tomo_queue.py/
test_tomo_queue_command_jobs.py that was blocking omny_e2e_tests collection
entirely, needed to actually verify this change's own test rewrites.

Co-Authored-By: Claude Sonnet 5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01NPZ9gzN2rt3wgiMnTKyrug
This commit is contained in:
x01dcandClaude Sonnet 5 committed 2026-10-01 15:11:49 +02:00
1 parent cf66e0d1ec
commit 437d9e7559
9 files changed
+208 -130

No files matched your search

@@ -175,9 +175,11 @@ class LamNI(TomoQueueMixin, LamNIAlignmentMixin, LamNIOpticsMixin, LamniGuiTools
}
# Named-action registry for tomo-queue command jobs -- identical to
# flomni's, since both actions (move, optimize_idgap) are beamline-wide,
# not flomni-specific hardware. See TomoQueueMixin._queue_action_move()/
# optimize_idgap() for the implementations.
# flomni's plain "move" action, since it's beamline-wide, not
# flomni-specific hardware. flomni also has a "move_foptz_with_shift"
# action for its own FZP optics-Z stage; lamni has no equivalent
# hardware, so it isn't mirrored here. See
# TomoQueueMixin._queue_action_move() for the implementation.
_TOMO_QUEUE_ACTIONS = {
"move": {
"func": "_queue_action_move",
@@ -185,21 +187,6 @@ class LamNI(TomoQueueMixin, LamNIAlignmentMixin, LamNIOpticsMixin, LamniGuiTools
"help": "Move device(s) to absolute position(s).",
"params": {"positions": {"type": "device_positions"}},
},
"optimize_idgap": {
"func": "optimize_idgap",
"idempotent": True,
"help": "Scan idgap over a search range and move to the peak. STUB: no-op.",
"params": {
"search_range": {
"type": "float",
"default": 0.5,
"min": 0.0,
"max": 2.0,
"unit": "mm",
"label": "Search range",
}
},
},
}
def _describe_tomo_job_line(self, job: dict) -> str:
@@ -586,6 +586,11 @@ class TomoQueueMixin:
if not isinstance(value, (int, float)) or isinstance(value, bool):
raise ValueError(f"{action_name}.{name} must be a number, got {value!r}.")
lo, hi = field.get("min"), field.get("max")
device_name = field.get("device_limits_from")
if device_name is not None:
device = self.device_manager.devices.get(device_name)
if device is not None:
lo, hi = device.limits
if lo is not None and value < lo:
raise ValueError(f"{action_name}.{name}={value} is below the minimum {lo}.")
if hi is not None and value > hi:
@@ -618,23 +623,6 @@ class TomoQueueMixin:
args += [dev[name], target]
umv(*args)
def optimize_idgap(self, search_range: float = 0.5) -> None:
"""Scan idgap over +/- search_range around its current position and
move to the peak of <signal not yet decided>.
STUB -- not implemented. The peak-detection signal/device is not yet
decided (see flomni's AI_docs/TOMO_QUEUE_COMMAND_JOBS_PLAN.md section
10). This is the ONLY place that needs to change to make this a real
action: the registry entry, resume policy, and (eventually) GUI
wiring already treat it as a normal idempotent compound action, so
dropping in the real scan-and-move-to-peak logic here (or swapping
which device/signal it reads) is a self-contained change.
"""
print(
f"optimize_idgap(search_range={search_range}): not yet implemented -- "
"peak-detection signal not yet decided. No-op."
)
def _describe_tomo_queue_actions(self) -> dict:
"""Serializable description of the action registry + move allow-list.
@@ -41,8 +41,10 @@ support on top of it:
registry `Flomni.__init__` publishes) so its action dropdown and
per-action fields always match whatever's actually in `flomni.py` — no
hardcoded mirror. Lets you add steps in order (action + its own fields
only — a device dropdown and target for `move`, a bounded spinbox for
`optimize_idgap`'s `search_range`), remove a step, reorder steps within
only — a device dropdown and target for `move`, three bounded spinboxes
for `move_foptz_with_shift`'s `foptz_target`/`manual_shift_x`/
`manual_shift_y`, the first of which ranges over `foptz`'s live device
limits rather than a static number), remove a step, reorder steps within
the job (move up/down), and set a label + idempotent flag (defaults to
"every step idempotent" ANDed, same as the CLI, until you touch the
checkbox yourself). On confirm it produces a job dict byte-for-byte
@@ -50,7 +52,8 @@ support on top of it:
2. **Command-job-aware queue table.** `TomoQueueDialog`'s table gained a
"Details" column; a command job shows `CMD` in the Type column and its
step summary + idempotency marker in Details (e.g.
`move{'positions': {'mokev': 6.2}} > optimize_idgap{'search_range': 0.5} [idem]`
`move{'positions': {'ccm_energy': 6.2}} > move_foptz_with_shift{'foptz_target':
15.0, 'manual_shift_x': 1.0, 'manual_shift_y': -2.0} [idem]`
— literally the same string `flomni.tomo_queue_show()` prints).
3. **Self-describing tomo jobs.** "Add current params to queue" now stamps
`"kind": "tomo"` and a fresh `"id"` on the job it writes, matching what
@@ -201,19 +204,22 @@ file is exactly how a regression sneaks in.
## 4. Test plan — new: command-job builder
6. Click **"Add command…"**. Confirm the dialog opens with an **action
dropdown showing `move` and `optimize_idgap`** (or whatever's currently
in `flomni.py`'s `_TOMO_QUEUE_ACTIONS` — if you've added a third action
since this was written, it should appear automatically with no GUI
change needed; that's the point of reading the live registry).
dropdown showing `move` and `move_foptz_with_shift`** (or whatever's
currently in `flomni.py`'s `_TOMO_QUEUE_ACTIONS` — if you've added a
third action since this was written, it should appear automatically with
no GUI change needed; that's the point of reading the live registry).
7. With `move` selected, confirm the field area shows a **device dropdown**
(should list `mokev`, `idgap`, `ftray` per the current allow-list) and a
**numeric target field**. Pick a device, set a target, click **"Add
step"**. Confirm a line appears in the step list above, formatted like
`move{'positions': {'mokev': 6.2}}`.
8. Switch the action dropdown to `optimize_idgap`. Confirm the field area
now shows a single **search_range** spinbox pre-filled with `0.5`,
range-limited (try typing outside 0–2, confirm the widget clamps it
rather than accepting it). Click **"Add step"**.
(should list `ccm_energy`, `idgap` per the current allow-list -- `foptz`
is deliberately absent, see point 8) and a **numeric target field**. Pick
a device, set a target, click **"Add step"**. Confirm a line appears in
the step list above, formatted like `move{'positions': {'ccm_energy': 6.2}}`.
8. Switch the action dropdown to `move_foptz_with_shift`. Confirm the field
area now shows three spinboxes: **foptz target (absolute)**,
**Compensating manual_shift_x**, **Compensating manual_shift_y**. Confirm
the foptz-target spinbox's range reflects `foptz`'s *live* device limits
(try typing a value past the real limit, confirm the widget clamps it
rather than accepting it -- this is the live-limit lookup, not a static
number). Fill in all three and click **"Add step"**.
9. Confirm the **idempotent checkbox auto-checked itself** after both adds
(both current actions default to idempotent=True). Uncheck it by hand,
then add another step — confirm it **stays unchecked** (your override
@@ -250,9 +256,10 @@ file is exactly how a regression sneaks in.
## 6. Test plan — mixed queue, end to end
17. Build a small realistic sequence via the GUI only: a tomo job ("Add
current params to queue"), then a command job (e.g. a `move` on
`ftray` if you're on the sim — `mokev`/`idgap` only exist on the real
front-end, see plan section 3.1), then another tomo job.
current params to queue"), then a command job (e.g. a `move_foptz_with_shift`
step if you're on the sim — `ccm_energy`/`idgap` only exist on the real
front-end, `foptz` is the only queue-movable device in the sim config),
then another tomo job.
18. From the CLI, run `flomni.tomo_queue_execute()`. Confirm: the tomo job
runs, the command job actually moves the device (check
`dev.<device>.readback.get()` before/after), and the second tomo job
@@ -264,8 +264,8 @@ interruptible mid-scan for the crash-resume tests (§4, B5).
10. **✅ Add-time validation rejects bad steps, and rejects them without landing
anything in the queue.** `test_add_command_validation_errors`. Unknown action,
a device not on the `move` allow-list, an out-of-range/wrong-type
`optimize_idgap.search_range`, and an unknown kwarg are each rejected with the
expected `ValueError`; afterward the queue is confirmed still empty.
`move_foptz_with_shift.foptz_target`, and an unknown kwarg are each rejected
with the expected `ValueError`; afterward the queue is confirmed still empty.
11. **✅ Execution-time allow-list re-check is the real safety boundary.**
`test_move_execution_time_allowlist_recheck`. A job with a disallowed device
appended directly (bypassing `tomo_queue_add_command`) is refused by
@@ -278,10 +278,13 @@ interruptible mid-scan for the crash-resume tests (§4, B5).
`test_show_and_webpage_survive_command_job`. A command job (no `params` key)
renders its step summary + idempotency marker in both, instead of `KeyError`ing.
14. **✅ Back-compat + real dispatch, combined.**
`test_legacy_job_no_kind_defaults_to_tomo_and_command_moves_ftray`. A `kind`-less
`test_legacy_job_no_kind_defaults_to_tomo_and_command_moves_foptz`. A `kind`-less
legacy queue entry still runs as a tomogram; in the same mixed-queue run, a
command job actually moves a device (`ftray`, the sim-testable stand-in for
`mokev`/`idgap` — see the plan doc §3.1) via the real `move` action path.
command job actually moves `foptz` and sets the compensating
`manual_shift_x`/`manual_shift_y` via the real `move_foptz_with_shift` action
path (the only sim-movable path now that `ftray` is no longer in the move
allow-list; `foptz` is also the real device this action exists for, unlike
`ftray`'s old sim-only stand-in role).
15. **✅ Non-idempotent resume prompts exactly once, and respects "no".**
`test_non_idempotent_command_resume_prompts`. A command job flagged
`idempotent=False`, left `incomplete`, triggers the resume prompt (distinct from
@@ -294,13 +297,14 @@ interruptible mid-scan for the crash-resume tests (§4, B5).
16. One queued tomogram end-to-end on the real instrument, to confirm the sim's
timing/progress semantics match reality closely enough that resume lands on the
right angle.
17. `move` on `mokev`/`idgap` end-to-end — neither device is in the flomni sim
config (see the plan doc §3.1), so Step 2's `move` action itself has only been
sim-exercised via `ftray`, never against the real devices it's actually for.
17. `move` on `ccm_energy`/`idgap` end-to-end — neither device is in the flomni
sim config, so Step 2's `move` action itself has only been sim-exercised via
add-time validation (which doesn't require real movement), never a real
execute() against these two devices.
**Gate:** A–D passed on the sim (16/16 tests total). **Steps 1 and 2 are verified
enough to build Step 3 (the GUI) on.** E remains open, and blocks calling `move` on
`mokev`/`idgap` production-ready.
**Gate:** A–D passed on the sim. **Steps 1 and 2 are verified enough to build
Step 3 (the GUI) on.** E remains open, and blocks calling `move` on
`ccm_energy`/`idgap` production-ready.
### A gotcha found testing Step 2 (not a Step-2 bug)
@@ -348,11 +352,14 @@ still just its design.
- **Registry entry = 4 things:** bound method name, `idempotent` default, help string,
and a **params schema** (possibly empty).
- **Result-dependent logic lives INSIDE a named action, never across queue steps.**
E.g. "scan idgap, move to peak" is ONE action (`optimize_idgap`) whose method does
the scan, finds the peak, and moves — *not* three queue steps. The queue therefore
never has to carry runtime values between steps; `kwargs` stays plain JSON. A
compound action may take parameters (e.g. `search_range`), declared in its params
schema like any other.
E.g. a hypothetical "scan idgap, move to peak" action would be ONE action whose
method does the scan, finds the peak, and moves — *not* three queue steps. The
queue therefore never has to carry runtime values between steps; `kwargs` stays
plain JSON. The same "bundle, don't split" idea also applies to a compound action
with no runtime-dependent result but a correctness-critical pairing of writes --
`move_foptz_with_shift` (moves `foptz` + sets the compensating
`manual_shift_x`/`manual_shift_y` atomically, deliberately not reachable as two
separate steps) is the concrete example currently in the registry.
- **Publish the registry via a global var** (`tomo_queue_actions`) on `Flomni.__init__`
so the GUI builds its dropdowns/forms from it — no hardcoded mirror to drift (the
`QUEUE_PARAM_NAMES` mirror between `flomni.py` and `tomo_params.py` already has this
@@ -399,9 +406,10 @@ Guarded against command jobs (`job.get("kind", "tomo") == "tomo"` before touchin
## 6. Open questions for Mirko — all resolved
- **The final action set and device allow-list**, see `TOMO_QUEUE_COMMAND_JOBS_PLAN.md`
§10 — `move` + `optimize_idgap` only, `_TOMO_QUEUE_MOVE_DEVICES = {mokev, idgap}`.
One detail (`optimize_idgap`'s peak-detection signal) deferred to the
implementation session, not blocking.
§10 — originally `move` + `optimize_idgap`, `_TOMO_QUEUE_MOVE_DEVICES = {mokev,
idgap}`. Superseded since: `optimize_idgap` was removed (never commissioned, stayed
a no-op stub), `mokev` was fixed to the real device name `ccm_energy`, and a new
`move_foptz_with_shift` action was added (see §4D10/§4D14 above for its tests).
- **`_TOMO_QUEUE_PARAM_NAMES` renamed to `_TOMO_SCAN_PARAM_NAMES`** (it reads live
properties, not queue state — the old name was misleading). Done in `flomni.py`;
the `tomo_params.py` comment referencing it by name was updated too (the
@@ -4455,14 +4455,15 @@ class Flomni(
# Curated allow-list for the "move" queue action -- NOT "any device in
# dev". Only devices actually reconfigured between tomograms belong here;
# this is the difference between a reconfiguration tool and a remote
# control for the whole endstation. mokev/idgap are real front-end
# devices (bl_frontend.yaml, not the flomni sim config); ftray (sample
# transfer tray, harmless to move, untouched by tomo-scan code) is
# included so the move action has a sim-testable target.
# control for the whole endstation. ccm_energy/idgap are real front-end
# devices (bl_frontend.yaml/bl_optics_hutch.yaml, not the flomni sim
# config). foptz is deliberately NOT here -- it has its own dedicated
# "move_foptz_with_shift" action below so the compensating
# manual_shift_x/y correction can never be forgotten (see that action's
# docstring).
_TOMO_QUEUE_MOVE_DEVICES = {
"mokev": {"label": "Energy", "unit": "keV"},
"ccm_energy": {"label": "Energy", "unit": "keV"},
"idgap": {"label": "Undulator gap", "unit": "mm"},
"ftray": {"label": "Sample transfer tray (sim-testable)", "unit": "mm"},
}
# Named-action registry for tomo-queue command jobs. Each entry maps a
@@ -4484,23 +4485,83 @@ class Flomni(
"help": "Move device(s) to absolute position(s).",
"params": {"positions": {"type": "device_positions"}},
},
"optimize_idgap": {
"func": "optimize_idgap",
"idempotent": True,
"help": "Scan idgap over a search range and move to the peak. STUB: no-op.",
"move_foptz_with_shift": {
"func": "_queue_action_move_foptz_with_shift",
"idempotent": True, # absolute move + absolute param set, re-runs harmlessly
"help": (
"Move foptz (FZP optics-Z) to an absolute target and set the "
"compensating manual_shift_x/manual_shift_y in the same step -- "
"changing foptz shifts the aligned scan center, so this is the "
"only way to queue a foptz move."
),
"params": {
"search_range": {
"foptz_target": {
"type": "float",
"default": 0.5,
"default": 17.0,
"min": 0.0,
"max": 2.0,
"max": 27.0,
"unit": "mm",
"label": "Search range",
}
"label": "foptz target (absolute)",
# Live ophyd limits (dev.foptz.limits) are checked at add
# time (CommandJobBuilderDialog._build_field_widget /
# TomoQueueMixin._validate_action_kwargs) and again at
# execution time (_queue_action_move_foptz_with_shift) --
# min/max above are only the static fallback if the live
# device can't be reached.
"device_limits_from": "foptz",
},
"manual_shift_x": {
"type": "float",
"default": 0.0,
"min": -10.0,
"max": 10.0,
"unit": "um",
"label": "Compensating manual_shift_x",
},
"manual_shift_y": {
"type": "float",
"default": 0.0,
"min": -50.0,
"max": 50.0,
"unit": "um",
"label": "Compensating manual_shift_y",
},
},
},
}
def _queue_action_move_foptz_with_shift(
self, foptz_target: float, manual_shift_x: float, manual_shift_y: float
) -> None:
"""Move foptz and set the compensating manual_shift_x/y together.
foptz changes the FZP-to-sample distance, which shifts the aligned
scan center (interferometer mirrors aren't perfectly parallel to the
beam) -- bundling the compensating shift into this one action is
deliberate: foptz is NOT in _TOMO_QUEUE_MOVE_DEVICES, so this is the
only way to queue a foptz move, precisely so the shift can't be
forgotten.
FUTURE: manual_shift_x/y could be auto-suggested from a calibration
proportional to the change in foptz, instead of typed in by hand --
deliberately out of scope here.
Re-checks foptz's live limits at execution time too (not just add
time, see TomoQueueMixin._validate_action_kwargs) -- the actual
safety boundary, since a value could have arrived via the CLI,
bypassing the add-time check, or the limits could have changed
since the job was queued.
"""
lo, hi = dev.foptz.limits
if not (lo <= foptz_target <= hi):
raise ValueError(
f"Queue move_foptz_with_shift: foptz_target={foptz_target} is "
f"outside foptz's current limits ({lo}, {hi})."
)
umv(dev.foptz, foptz_target)
self.manual_shift_x = manual_shift_x
self.manual_shift_y = manual_shift_y
def _describe_tomo_job_line(self, job: dict) -> str:
"""flomni's tomo_queue_show() one-line summary: fov/type/step/ctime/range.
@@ -2191,7 +2191,16 @@ class CommandJobBuilderDialog(QDialog):
if field_type == "float":
w = QDoubleSpinBox()
w.setDecimals(4)
w.setRange(field.get("min", -1e9), field.get("max", 1e9))
lo, hi = field.get("min", -1e9), field.get("max", 1e9)
device_name = field.get("device_limits_from")
if device_name is not None:
device = self._client.device_manager.devices.get(device_name)
if device is not None:
try:
lo, hi = device.limits
except Exception:
pass # fall back to the static schema min/max above
w.setRange(lo, hi)
w.setValue(field.get("default", 0.0))
return w
if field_type == "device_positions":
+1 -1
View File
@@ -25,7 +25,7 @@ from csaxs_bec.bec_ipython_client.plugins.flomni import flomni as flomni_module
from csaxs_bec.bec_ipython_client.plugins.flomni.flomni_webpage_generator import (
FlomniWebpageGenerator,
)
from csaxs_bec.bec_ipython_client.plugins.omny.omny_general_tools import OMNYTools
from csaxs_bec.bec_ipython_client.plugins.OMNY_shared.omny_general_tools import OMNYTools
class _NoOpSciLogMessage:
+1 -1
View File
@@ -65,7 +65,7 @@ def test_empty_and_all_done_queue_are_noops(flomni_sim, monkeypatch, capsys):
flomni = flomni_sim
prompt_calls = []
from csaxs_bec.bec_ipython_client.plugins.omny.omny_general_tools import OMNYTools
from csaxs_bec.bec_ipython_client.plugins.OMNY_shared.omny_general_tools import OMNYTools
monkeypatch.setattr(
OMNYTools, "yesno", lambda self, *a, **k: prompt_calls.append((a, k)) or True
+58 -40
View File
@@ -1,10 +1,13 @@
"""Step 2 -- command jobs. See AI_docs/TOMO_QUEUE_COMMAND_JOBS_PLAN.md.
mokev/idgap (the real curated move targets) aren't in the flomni sim config
(simulated_flomni.yaml) -- only on the real front end (bl_frontend.yaml), so
the real device-moving path here is exercised against `ftray` (sample
transfer tray), added to _TOMO_QUEUE_MOVE_DEVICES specifically for this
sim-testability, per TOMO_QUEUE_TESTING.md.
ccm_energy/idgap (the plain "move" allow-list) aren't in the flomni sim
config (simulated_flomni.yaml) -- only on the real front end
(bl_frontend.yaml/bl_optics_hutch.yaml). Tests that only exercise add-time
validation/rendering (no tomo_queue_execute()) reference idgap for that
reason, while tests that actually execute the queue use the dedicated
"move_foptz_with_shift" action against foptz, which IS in the sim config
and is the only queue-movable path for foptz (it is deliberately excluded
from the plain "move" allow-list, see flomni.py's _TOMO_QUEUE_MOVE_DEVICES).
"""
import datetime
@@ -13,8 +16,8 @@ import uuid
import pytest
from _queue_helpers import FAST_STEPSIZE, add_short_job
FTRAY_A = -50.0
FTRAY_B = -10.0
FOPTZ_A = 15.0
FOPTZ_B = 5.0
def test_add_command_validation_errors(flomni_sim):
@@ -28,12 +31,18 @@ def test_add_command_validation_errors(flomni_sim):
with pytest.raises(ValueError, match="above the maximum"):
flomni.tomo_queue_add_command(
{"action": "optimize_idgap", "kwargs": {"search_range": 99.0}}
{
"action": "move_foptz_with_shift",
"kwargs": {"foptz_target": 99.0, "manual_shift_x": 0.0, "manual_shift_y": 0.0},
}
)
with pytest.raises(ValueError, match="must be a number"):
flomni.tomo_queue_add_command(
{"action": "optimize_idgap", "kwargs": {"search_range": "wide"}}
{
"action": "move_foptz_with_shift",
"kwargs": {"foptz_target": "far", "manual_shift_x": 0.0, "manual_shift_y": 0.0},
}
)
with pytest.raises(ValueError, match="has no parameter"):
@@ -74,23 +83,25 @@ def test_move_execution_time_allowlist_recheck(flomni_sim):
def test_tomo_queue_actions_global_var_matches_registry(flomni_sim):
flomni = flomni_sim
published = flomni.client.get_global_var("tomo_queue_actions")
assert set(published["actions"]) == {"move", "optimize_idgap"}
assert published["move_devices"]["ftray"]["unit"] == "mm"
assert set(published["actions"]) == {"move", "move_foptz_with_shift"}
assert published["move_devices"]["idgap"]["unit"] == "mm"
assert published["actions"]["move"]["idempotent"] is True
assert published["actions"]["optimize_idgap"]["params"]["search_range"]["default"] == 0.5
assert (
published["actions"]["move_foptz_with_shift"]["params"]["foptz_target"]["default"] == 17.0
)
def test_show_and_webpage_survive_command_job(flomni_sim, capsys):
flomni = flomni_sim
flomni.tomo_queue_add_command(
{"action": "move", "kwargs": {"positions": {"ftray": FTRAY_A}}},
label="move ftray",
{"action": "move", "kwargs": {"positions": {"idgap": 6.0}}},
label="move idgap",
idempotent=True,
)
capsys.readouterr()
jobs = flomni.tomo_queue_show()
out = capsys.readouterr().out
assert "CMD move{'positions': {'ftray': -50.0}}" in out
assert "CMD move{'positions': {'idgap': 6.0}}" in out
assert "[idem]" in out
assert jobs[0]["kind"] == "command"
@@ -100,7 +111,7 @@ def test_show_and_webpage_survive_command_job(flomni_sim, capsys):
assert "isCommand" in html # the rendering branch exists; full JS isn't executed here
def test_legacy_job_no_kind_defaults_to_tomo_and_command_moves_ftray(flomni_sim):
def test_legacy_job_no_kind_defaults_to_tomo_and_command_moves_foptz(flomni_sim):
"""B-style back-compat + the real command-job execution path, combined
into one mixed-queue run to keep the (only) real scan in this file to
one job.
@@ -109,12 +120,12 @@ def test_legacy_job_no_kind_defaults_to_tomo_and_command_moves_ftray(flomni_sim)
bec = flomni.client
# Start from a known position, not whatever a previous test run happened
# to leave ftray at -- ftray's position is real device-server state that
# to leave foptz at -- foptz's position is real device-server state that
# persists across separate pytest invocations, unlike the tomo_queue/
# tomo_progress global vars the flomni_sim fixture snapshots/restores.
flomni._queue_action_move({"ftray": FTRAY_B})
ftray_before = bec.device_manager.devices.ftray.readback.get()
assert ftray_before == pytest.approx(FTRAY_B)
flomni._queue_action_move_foptz_with_shift(FOPTZ_B, 0.0, 0.0)
foptz_before = bec.device_manager.devices.foptz.readback.get()
assert foptz_before == pytest.approx(FOPTZ_B)
add_short_job(flomni, "legacy_tomo", FAST_STEPSIZE)
# Simulate a queue entry written before the "kind" field existed.
@@ -123,41 +134,48 @@ def test_legacy_job_no_kind_defaults_to_tomo_and_command_moves_ftray(flomni_sim)
bec.set_global_var("tomo_queue", jobs)
flomni.tomo_queue_add_command(
{"action": "move", "kwargs": {"positions": {"ftray": FTRAY_A}}}, label="move ftray"
)
flomni.tomo_queue_add_command(
{"action": "optimize_idgap", "kwargs": {"search_range": 0.5}}, label="idgap stub"
{
"action": "move_foptz_with_shift",
"kwargs": {"foptz_target": FOPTZ_A, "manual_shift_x": 1.0, "manual_shift_y": -2.0},
},
label="move foptz",
)
flomni.tomo_queue_execute()
jobs = flomni._tomo_queue_proxy.as_list()
assert [j["status"] for j in jobs] == ["done", "done", "done"]
assert [j["status"] for j in jobs] == ["done", "done"]
# Still no "kind" key -- proves it ran (and got marked done) via the
# back-compat default, not that execute() silently added one.
assert "kind" not in jobs[0]
assert "params" in jobs[0]
assert jobs[1]["kind"] == "command"
ftray_after = bec.device_manager.devices.ftray.readback.get()
assert ftray_after == pytest.approx(FTRAY_A)
assert ftray_after != pytest.approx(ftray_before)
foptz_after = bec.device_manager.devices.foptz.readback.get()
assert foptz_after == pytest.approx(FOPTZ_A)
assert foptz_after != pytest.approx(foptz_before)
assert flomni.manual_shift_x == pytest.approx(1.0)
assert flomni.manual_shift_y == pytest.approx(-2.0)
def test_non_idempotent_command_resume_prompts(flomni_sim, monkeypatch):
flomni = flomni_sim
bec = flomni.client
ftray = bec.device_manager.devices.ftray
foptz = bec.device_manager.devices.foptz
# Use the same real move path the "move" action itself uses, rather than
# the device's underlying ophyd object directly (ftray's BEC wrapper is
# a Positioner here, unlike fsamx's DeviceBase -- no .obj to reach for).
flomni._queue_action_move({"ftray": FTRAY_B})
start = ftray.readback.get()
assert start == pytest.approx(FTRAY_B)
# Use the same real move path the "move_foptz_with_shift" action itself
# uses, rather than the device's underlying ophyd object directly
# (foptz's BEC wrapper is a Positioner here, unlike fsamx's DeviceBase --
# no .obj to reach for).
flomni._queue_action_move_foptz_with_shift(FOPTZ_B, 0.0, 0.0)
start = foptz.readback.get()
assert start == pytest.approx(FOPTZ_B)
index = flomni.tomo_queue_add_command(
{"action": "move", "kwargs": {"positions": {"ftray": FTRAY_A}}},
{
"action": "move_foptz_with_shift",
"kwargs": {"foptz_target": FOPTZ_A, "manual_shift_x": 0.0, "manual_shift_y": 0.0},
},
label="non-idempotent move",
idempotent=False,
)
@@ -165,7 +183,7 @@ def test_non_idempotent_command_resume_prompts(flomni_sim, monkeypatch):
flomni._tomo_queue_proxy.update_by_id(job_id, status="incomplete")
prompts = []
from csaxs_bec.bec_ipython_client.plugins.omny.omny_general_tools import OMNYTools
from csaxs_bec.bec_ipython_client.plugins.OMNY_shared.omny_general_tools import OMNYTools
def _yesno(resume_answer):
# tomo_queue_execute()'s own top-level "OK to start?" confirmation
@@ -180,11 +198,11 @@ def test_non_idempotent_command_resume_prompts(flomni_sim, monkeypatch):
return _fn
# "no" -> must not re-run: ftray stays put, job still ends up "done".
# "no" -> must not re-run: foptz stays put, job still ends up "done".
monkeypatch.setattr(OMNYTools, "yesno", _yesno(False))
flomni.tomo_queue_execute()
assert len(prompts) == 1
assert ftray.readback.get() == pytest.approx(start)
assert foptz.readback.get() == pytest.approx(start)
assert flomni._tomo_queue_proxy.as_list()[0]["status"] == "done"
# Re-arm as incomplete and try again with "yes" -> must re-run.
@@ -192,5 +210,5 @@ def test_non_idempotent_command_resume_prompts(flomni_sim, monkeypatch):
monkeypatch.setattr(OMNYTools, "yesno", _yesno(True))
flomni.tomo_queue_execute()
assert len(prompts) == 2
assert ftray.readback.get() == pytest.approx(FTRAY_A)
assert foptz.readback.get() == pytest.approx(FOPTZ_A)