## Operating principle
|
|
Single shot diagnostic to measure relative pulse energy and beam position. Back-scattered X-rays from a thin target (typically 50 nm silicon nitride) are measured by four photodiodes positioned above, below, left and right of the X-ray beam.
|
|
|
|
Position data maintains FEL pointing, while single-shot pulse energy measurements are used to normalise experiment data. There is one PBPS in the front-end and two further devices per end station.
|
|
|
|
## How to use
|
|
|
|
1. Open photon diagnostics [web app](https://sf-photodiag.psi.ch/app)
|
|
2. Under Position/calibration tabs,s elect device from drop down menu 1
|
|
3. Select target from drop down menu 2 - default target is 50nm Si3N4
|
|
4. Press calibrate
|
|
<img src="https://docs.google.com/drawings/d/e/2PACX-1vRGynTQ54ugugA_bM9pYc9eZDlet3Z9MSZWn17wJNwAEM2ORiJGoL8O61frpLBVRpFKlzJ7M05a5Mb1/pub?w=1245&h=1039">
|
|
|
|
The calibration procedure normalizes the response of the 4 diodes, then scans the device horizontally and vertically over the FEL beam to calibrate the position response.
|
|
|
|
Once completed (approx 1 minute) the plots are updated with the calibration response. Plotted on the x axis is the device motor position (horizontal - left plot and vertical - right plot) and on the y axis is plotted the normalized horizontal or vertical diode pair response. For a good setup the plot should be a straight line through the 3 data points.
|
|
|
|
If you are happy with the calibration, press the "Push results/elog" button where the new calibration data is sent to the required data processing pipelines. After an automated restart (~30 seconds) both PV and BS position and intensity data will be calculated with the new calibration. An elog entry is made detailing the calibration data.
|
|
|
|
## Error
|
|
**Soft limits**
|
|
Error message: `ERROR: Motor position outside soft limits`
|
|
Problem is a soft limit on either the device Motor X or Y is blocking the scan motion
|
|
1. Select all from drop down menu for motor X and motor Y of PBPS GUI (shown below)
|
|
2. Change motor soft limits to 1 and -1 for both motors
|
|
3. Retry calibration
|
|
<img src="https://docs.google.com/drawings/d/e/2PACX-1vSga-Zluo4b0isNs92C5MXfawMueNVk3ideRRDA_ZGQmD05Jn4noTS_pOl0OHiOyjS31OhLTAelFkYn/pub?w=728&h=686"> |