Update PBPS calibration guide
This commit is contained in:
parent
de3d052e8c
commit
42876cf108
@ -15,4 +15,13 @@ The calibration procedure normalizes the response of the 4 diodes, then scans th
|
||||
|
||||
Once completed (approx 1 minute) the plots are updated with the calibration response. Plotted on the x axis is the device motor position (horizontal - left plot and vertical - right plot) and on the y axis is plotted the normalized horizontal or vertical diode pair response. For a good setup the plot should be a straight line through the 3 data points.
|
||||
|
||||
If you are happy with the calibration, press the "Push results/elog" button where the new calibration data is sent to the required data processing pipelines. After an automated restart (~30 seconds) both PV and BS position and intensity data will be calculated with the new calibration. An elog entry is made detailing the calibration data.
|
||||
If you are happy with the calibration, press the "Push results/elog" button where the new calibration data is sent to the required data processing pipelines. After an automated restart (~30 seconds) both PV and BS position and intensity data will be calculated with the new calibration. An elog entry is made detailing the calibration data.
|
||||
|
||||
## Error
|
||||
**Soft limits**
|
||||
Error message: `ERROR: Motor position outside soft limits`
|
||||
Problem is a soft limit on either the device Motor X or Y is blocking the scan motion
|
||||
1. Select all from drop down menu for motor X and motor Y of PBPS GUI (shown below)
|
||||
2. Change motor soft limits to 1 and -1 for both motors
|
||||
3. Retry calibration
|
||||
<img src="https://docs.google.com/drawings/d/e/2PACX-1vSga-Zluo4b0isNs92C5MXfawMueNVk3ideRRDA_ZGQmD05Jn4noTS_pOl0OHiOyjS31OhLTAelFkYn/pub?w=728&h=686">
|