d32c8d526bf956ddaa952e7a6dc13822f45292fd
Both directions of the tilt now go through the same restraint, with a budget each in shared direct-beam pixels; zero means free. Swapping which component is held, or holding both, is a change to two constants. The perpendicular budget is zero today, which is the arrangement the measurements support: it is the component whose conditioning tracks the 2theta the fit reaches, while the parallel one's does not move with 2theta at all - a conditioning number that ignores the data is the prior talking. This is a clarity change, not a precision one, and the record should not read otherwise: the standing 'freeing the tilt inflates both beam components' cost is the perpendicular component's doing, not the parallel one's, so restraining the parallel one recovers essentially none of it. What it buys is that a beam-centre error is no longer reported as an angle.
Jungfraujoch
Application to receive data from the PSI JUNGFRAU and EIGER detectors.
All documentation is now placed in docs/ subdirectory and for the current version hosted on Jungfraujoch Read The Docs page.
Languages
C++
76.7%
HTML
7.2%
C
5.6%
TypeScript
4%
Cuda
2.1%
Other
4.3%