c095b19f3afc254cd75c8f6261d64bd842a550c2
The XDS references of aspirin (20, 25 keV), citric acid, HEPES and YAG were made with OVERLOAD=65534, half of the EIGER2's saturation value (133202). CORRECT ignored 13-313 reflections per set whose peak pixel lay between the two - the strongest, most extinguished reflections - so the references described a data set without them. Re-ran only the CORRECT step on the unchanged INTEGRATE.HKL with every input parameter reconstructed from the old CORRECT.LP and OVERLOAD=133201 (saturation - 1, as the other in-house sets already use); a rerun at 65534 reproduced the old CORRECT.LP. The old outputs are kept in xds_overload_half/ beside each data set. `refs --write` picked up the new values. Co-Authored-By: Claude Opus 5.5 (1M context) <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01K5K8jvPPbmCrbqnWkddTuB