bf6efc7fe9d72d8773e51083bef2ca6d935a2b13
Bumped the default signal-box radius to r1=4 (r2=6, r3=8): on the lysozyme jet (1% DMM) it lifts CCref 50.1->52.2% and CC1/2 90.8->92.5% (its broadened spots spill past a radius-3 box), is neutral on the mono crystal and on the classical integrator. Added 'jfjoch_process --integration-radius <r1|r1,r2,r3>' (a single value derives r2=r1+2, r3=r1+4) and wired PixelRefine's shoebox radius to BraggIntegrationSettings::GetR1() so it shares the classical integrator's knob. (Explored but rejected: an elliptical/anisotropic Term-2 profile. The jet's tangential spots are mildly anisotropic - axis ratio 1.15->1.44 low->high res, azimuthal mosaic, separate from the radial DMM bandwidth - but using the measured 2x2 covariance as the profile DEGRADED the jet, CC1/2 92.5->83.5: the tight measured width loses to a generous aperture, same lesson as the radius bump.) Co-Authored-By: Claude Opus 4.8 <noreply@anthropic.com>
Jungfraujoch
Application to receive data from the PSI JUNGFRAU and EIGER detectors.
All documentation is now placed in docs/ subdirectory and for the current version hosted on Jungfraujoch Read The Docs page.
Languages
C++
73.7%
C
8.8%
HTML
7.8%
TypeScript
4%
Tcl
3.3%
Other
2.3%