51e90f554956265f217ee7aac64632d67dbfff2f
The report invited the reader to check REFINED_DETECTOR_TILT against a powder calibration. Measured over two detectors' worth of crystals, that check misleads: the rotation fit does one outer round, so it leaves its starting value by only a small and crystal-dependent fraction of the distance to the calibrated value, and on the worse of the two detectors the per-crystal median sits an order of magnitude further from the powder answer than either method's uncertainty. A user comparing one run against their own calibration would conclude the calibration was wrong. What survives the aliasing is the direct beam, which is already printed beside it, and the MEDIAN of the tilt over several crystals on one detector - enough to show up a placeholder or a stale value in the file, not enough to replace a calibration. Say that, and say not to feed one run's value back into the instrument. The conditioning law (VIF = 4.40 tan(2theta_95)^-1.66) was tested as a per-run gate and does NOT order the errors - the crystals whose fit stays put are found in the best-conditioned band as often as the worst - so no threshold is added here. Co-Authored-By: Claude Opus 5 (1M context) <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01EFEJG6WBQv8th4UJFNe53N
Jungfraujoch
Application to receive data from the PSI JUNGFRAU and EIGER detectors.
All documentation is now placed in docs/ subdirectory and for the current version hosted on Jungfraujoch Read The Docs page.