3be79fb1b6acf485421bd14e384d451df7049039
The warning said the diffraction limit "runs from 3.34 to 2.45 A depending on direction" and stopped there, so a reader was told the crystal is anisotropic and given no direction to act on. It fires on 19 of 113 datasets in the battery. The eigenvectors are already measured and already written to the merged mmCIF as `_reflns.pdbx_aniso_B_tensor_eigenvector_N_ortho`; they had simply never reached the human report. The tensor is fitted on s = frac.mat * (h,k,l), so in that Cartesian frame a*, b*, c* are the rows of frac.mat and naming the axis is one dot product per eigenvector. The label is exact in every Laue class that has a free tensor direction except triclinic, and the cosine is printed so a loose fit shows as one. Co-Authored-By: Claude Opus 5 (1M context) <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01EFEJG6WBQv8th4UJFNe53N
Jungfraujoch
Application to receive data from the PSI JUNGFRAU and EIGER detectors.
All documentation is now placed in docs/ subdirectory and for the current version hosted on Jungfraujoch Read The Docs page.
Languages
C++
76.7%
HTML
7.2%
C
5.6%
TypeScript
4%
Cuda
2.1%
Other
4.3%