040810aa3d0d8ec91ef790ba37c2568f272ab3a7
Where the file's and the measured beam centre index lattices whose primitive volumes differ by an integer factor 2-4, the pooled validation evidence decided only when the measured centre held the LARGER cell. The other direction was left to the second pass on the assumption that it drops to the smaller cell by itself. It does so only when the pass-2 indexer's free beam refinement walks the whole centre error, and that rescue is chaotic: on 7ris a 0.1 % change in the pass-1 distance switched off a 6.5 px walk, the doubled c axis (380 A) won and the run failed. Both directions are now decided on the same measurement: pooled validation spots on each lattice against its own wrong-spindle null (ValidationEvidencePrefers). No new threshold. 7ris: 44.3 % at the file's centre vs 83.5 % at the measured one; the run adopts the measured centre and the 190 A cell (R_free 0.81 -> 0.20, 1.52 A). 8pqd (3x, measured centre holds the true cell): 16.6 % vs 21.0 %, adopted in pass 1 instead of being recovered by pass 2. Co-Authored-By: Claude Opus 5.5 (1M context) <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01D1G8gJVAy6gp1K5Dz3NE5C