Don't need to turn on multiple solenoids as one-way valve appears to be fitted
r3294 | dcl | 2011-11-29 18:51:35 +1100 (Tue, 29 Nov 2011) | 1 line
This commit is contained in:
@@ -407,8 +407,6 @@ SHF;' TURN SAMPLE STAGE MOTOR ON
|
||||
WT500;' WAIT FOR MOTOR TO STABILISE
|
||||
SB 10;'SAMPLE STAGE AIR PAD OFF-SOLENOID OFF
|
||||
CB 9;'SAMPLE STAGE AIR PAD ON-SOLENOID ON
|
||||
SB 12;'DETECTOR AIR PAD OFF-SOLENOID OFF
|
||||
CB 11;'DETECTOR AIR PAD ON-SOLENOID ON
|
||||
WT1000;
|
||||
RSPF=1
|
||||
ENDIF
|
||||
@@ -416,15 +414,8 @@ IF (REQF=0)&(RSPF=1)&(_BGF=0);'MOTION FINISHED
|
||||
WT1000;' WAIT FOR MOTOR TO SETTLE
|
||||
SB 9;'SAMPLE STAGE AIR PAD ON-SOLENOID OFF
|
||||
CB 10;'SAMPLE STAGE AIR PAD OFF-SOLENOID ON
|
||||
IF (REQH=0)&(RSPH=0);'DETECTOR IS NOT ON
|
||||
SB 11;' DETECTOR AIR PAD ON-SOLENOID OFF
|
||||
CB 12;' DETECTOR AIR PAD OFF-SOLENOID ON
|
||||
ENDIF
|
||||
WT500;'WAIT 1/2 SEC
|
||||
SB 10;'SAMPLE STAGE AIR PAD OFF-SOLENOID OFF
|
||||
IF (REQH=0)&(RSPH=0);'DETECTOR IS NOT ON
|
||||
SB 12;' DETECTOR AIR PAD OFF-SOLENOID OFF
|
||||
ENDIF
|
||||
MOF;' TURN SAMPLE STAGE MOTOR OFF
|
||||
RSPF=0
|
||||
ENDIF
|
||||
@@ -432,11 +423,11 @@ ENDIF
|
||||
'CONTROLS THE AIRPAD SOLENOID VALVES FOR DETECTOR STAGE
|
||||
' FOLLOW SHF/MOF COMMANDS ENTERED MANUALLY
|
||||
IF (_MOH=0)&(RSPH=0);'MOTOR IS ON BUT AIR IS OFF
|
||||
REQF=1
|
||||
REQH=1
|
||||
ENDIF
|
||||
IF (_MOH=1)&(RSPH=1);'MOTOR IS OFF BUT AIR IS ON
|
||||
SHF;' FOLLOWING STEPS MUST BE RUN BEFORE MOTOR CAN BE TURNED OFF
|
||||
REQF=0
|
||||
REQH=0
|
||||
ENDIF
|
||||
'
|
||||
IF (REQH=1)&(RSPH=0);'REQUESTING DETECTOR MOTOR BE MOVABLE
|
||||
@@ -449,12 +440,10 @@ RSPH=1
|
||||
ENDIF
|
||||
IF (REQH=0)&(RSPH=1)&(_BGH=0);'MOTION FINISHED
|
||||
WT1000;' WAIT FOR MOTOR TO SETTLE
|
||||
IF (REQF=0)&(RSPF=0);'SAMPLE STAGE IS NOT ON
|
||||
SB 11;'DETECTOR AIR PAD ON-SOLENOID OFF
|
||||
CB 12;'DETECTOR AIR PAD OFF-SOLENOID ON
|
||||
WT500;'WAIT 1/2 SEC
|
||||
SB 12;'DETECTOR AIR PAD OFF-SOLENOID OFF
|
||||
ENDIF
|
||||
MOH;' TURN MOTOR OFF
|
||||
RSPH=0
|
||||
ENDIF
|
||||
@@ -673,3 +662,5 @@ EN
|
||||
'101 SAFETY AND INTERLOCK SYSTEM HAS INITIATED A SLOWDOWN BY LIGHT CURTAIN
|
||||
'102 SAFETY AND INTERLOCK SYSTEM HAS INITIATED A SLOWDOWN BY LIGHT CURTAIN
|
||||
'
|
||||
|
||||
|
||||
|
||||
Reference in New Issue
Block a user