Get voltage settings as measured by Mark Lesha on 30/7/13 for SN003

This commit is contained in:
Ferdi Franceschini
2013-08-15 08:07:39 +10:00
parent 36caadd3af
commit dd0ee77a26

View File

@@ -15,15 +15,20 @@ fileeval $cfPath(hmm)/sct_orhvps_common.tcl
#dhv1 lower 19
#dhv1 upper 59
# 22/6/2011
# MJL 30/7/13 NOTE: Settings below reflect the manufacturer-recommanded voltages set at last service by Ordela.
# S/N 006 last serviced 23 Apr 2012, recommended voltage 2100V.
set dhv1Max_SN006 2405
set upper_SN006 55
set dhv1Max_SN003 2600
set upper_SN003 57
# S/N 003 last serviced 31 May 2011, recommended voltage 2300V.
# Observed a setting of 61 gives 2294V on the voltage monitor, implying a max voltage of 2370V.
set dhv1Max_SN003 2370
set upper_SN003 61
set dhvMax $dhv1Max_SN006
set dhvUpper $upper_SN006
# At 30/7/13, S/N 003 is installed on Quokka (S/N 006 electronics failed, being repaired at ANSTO)
set dhvMax $dhv1Max_SN003
set dhvUpper $upper_SN003
set host [dict get $::HISTMEM_HOSTPORT ORDELA HOST]
set port [dict get $::HISTMEM_HOSTPORT ORDELA PORT]