diff --git a/site_ansto/instrument/sans/config/hmm/ordela_detector_config.tcl b/site_ansto/instrument/sans/config/hmm/ordela_detector_config.tcl index 01d50e3e..c46e1d01 100644 --- a/site_ansto/instrument/sans/config/hmm/ordela_detector_config.tcl +++ b/site_ansto/instrument/sans/config/hmm/ordela_detector_config.tcl @@ -15,15 +15,20 @@ fileeval $cfPath(hmm)/sct_orhvps_common.tcl #dhv1 lower 19 #dhv1 upper 59 -# 22/6/2011 +# MJL 30/7/13 NOTE: Settings below reflect the manufacturer-recommanded voltages set at last service by Ordela. + +# S/N 006 last serviced 23 Apr 2012, recommended voltage 2100V. set dhv1Max_SN006 2405 set upper_SN006 55 -set dhv1Max_SN003 2600 -set upper_SN003 57 +# S/N 003 last serviced 31 May 2011, recommended voltage 2300V. +# Observed a setting of 61 gives 2294V on the voltage monitor, implying a max voltage of 2370V. +set dhv1Max_SN003 2370 +set upper_SN003 61 -set dhvMax $dhv1Max_SN006 -set dhvUpper $upper_SN006 +# At 30/7/13, S/N 003 is installed on Quokka (S/N 006 electronics failed, being repaired at ANSTO) +set dhvMax $dhv1Max_SN003 +set dhvUpper $upper_SN003 set host [dict get $::HISTMEM_HOSTPORT ORDELA HOST] set port [dict get $::HISTMEM_HOSTPORT ORDELA PORT]