@@ -62,7 +62,7 @@ _To bypass the fine alignment: `flomni.feye_out`_
1. `flomni.tomo_parameters()` Adjust the ptychographic scan parameters for performing an alignment scan. Typically FOVX = FOVX(Xrayeye)+20 mu, shell step = beamsize/2.5, number of projections and tomo mode are ignored in the alignment scans.
1. `flomni.tomo_alignment_scan()` perform the alignment scan. When the first scan is running, switch to a matlab session and run `SPEC_ptycho_align` again. Click left and right. The third click can define the height of the scan, but is not needed and ignored by default. The widest horizontal field of view will be printed at the end of the matlab session.
1. `flomni.tomo_alignment_scan()` perform the alignment scan. When the first scan is running, switch to a matlab session and run `BEC_ptycho_align` again. Click left and right. The third click can define the height of the scan, but is not needed and ignored by default. The widest horizontal field of view will be printed at the end of the matlab session.
1. `flomni.read_alignment_offset()` Load alignment parameters calculated in matlab.
### Tomographic Measurement