Design pipeline:
|Set|Get|Check|
|---|---|----|
|defocus to 0 | | |
|beam energy | contrast transfer function | |
|probe semi-angle | contrast transfer function | spatial frequencies in sample |
|magnification | field of view | |
|scanning matrix | sampling | for atomic resolution around 0.2 - 0.4 A|
|camera length | detector cover| just over 1 alfa, check for recommended |
|dwell time | electron dose | consider binning and maximal detector frame rate |
|probe current | electron dose | compare with sample hardness, is it reasonable for given resolution?|
||Set|Get|Check|
|---|---|---|----|
|1.|defocus to 0 | | |
|2.|beam energy | contrast transfer function | |
|3.|probe semi-angle | contrast transfer function | spatial frequencies in sample |
|4.|magnification | field of view | |
|5.|scanning matrix | sampling | for atomic resolution around 0.2 - 0.4 A|
|6.|camera length | detector cover| just over 1 alfa, check for recommended |
|7.|dwell time | electron dose | consider binning and maximal detector frame rate |
|8.|probe current | electron dose | compare with sample hardness, is it reasonable for given resolution?|
![instructionsssb](uploads/6d89164fc1d41ce256487045f767a1d7/instructionsssb.png)
# Iterative reconstruction