Update Instructions
@ -6,16 +6,19 @@
|
||||
|
||||
Design pipeline:
|
||||
|
||||
|Set|Get|Check|
|
||||
|---|---|----|
|
||||
|defocus to 0 | | |
|
||||
|beam energy | contrast transfer function | |
|
||||
|probe semi-angle | contrast transfer function | spatial frequencies in sample |
|
||||
|magnification | field of view | |
|
||||
|scanning matrix | sampling | for atomic resolution around 0.2 - 0.4 A|
|
||||
|camera length | detector cover| just over 1 alfa, check for recommended |
|
||||
|dwell time | electron dose | consider binning and maximal detector frame rate |
|
||||
|probe current | electron dose | compare with sample hardness, is it reasonable for given resolution?|
|
||||
||Set|Get|Check|
|
||||
|---|---|---|----|
|
||||
|1.|defocus to 0 | | |
|
||||
|2.|beam energy | contrast transfer function | |
|
||||
|3.|probe semi-angle | contrast transfer function | spatial frequencies in sample |
|
||||
|4.|magnification | field of view | |
|
||||
|5.|scanning matrix | sampling | for atomic resolution around 0.2 - 0.4 A|
|
||||
|6.|camera length | detector cover| just over 1 alfa, check for recommended |
|
||||
|7.|dwell time | electron dose | consider binning and maximal detector frame rate |
|
||||
|8.|probe current | electron dose | compare with sample hardness, is it reasonable for given resolution?|
|
||||
|
||||
|
||||

|
||||
|
||||
|
||||
# Iterative reconstruction
|
||||
|