Update Controls

2024-03-15 17:18:37 +01:00
parent 8bfec1c7d8
commit 37c80647e0

@ -18,11 +18,11 @@ current
- **Use PAAR** Pumping Aperture Angular Restriction, build-in aperture may limit maximal scattering angles for shortest camera lengths
- **Binning** reduction in pixel array size usually increases maximum frame rate
##### CL guide - build in checks for proper camera length selections for iterative reconstruction
- low angle approximation - scattering angle < 10deg (unequally sampled pixel around optical axis and around detector edge)
- ptycho pixel < step size - set step size is smaller than final pixel in reconstructed image - loose of resolution
- beam fits in probe window - probe window is capable to accommodate for defocused probe
- detector cover < 1α - no reconstruction method for such a dataset
- detector cover > 6α - usually no scattering beyond 5α
- Low angle approximation - scattering angle < 10deg (unequally sampled pixel around optical axis and around detector edge)
- Ptycho pixel < step size - set step size is smaller than final pixel in reconstructed image - loose of resolution
- Beam fits in probe window - probe window is capable to accommodate for defocused probe
- Detector cover < 1α - no reconstruction method for such a dataset
- Detector cover > 6α - usually no scattering beyond 5α
Always carefully check maximal detector frame rate, needed binning and set dwell time!!