#Parameters
|
||||
|
||||
B1 = 14 # starting mag. field in Amps
|
||||
B2 = -14# final mag. field in Amps
|
||||
BSTEP = 0.2 # step size mag. field in Amps
|
||||
ENERGIES = (770.0, 779.3) #list of energies in eV
|
||||
MODE = 1 #polarization (CIRC+ -> 1 or CIRC- -> 2)
|
||||
OFFSET1 = -1.90 #ID1 offset
|
||||
OFFSET2 = -1.90 #ID2 offset
|
||||
FIELD_PRECISION = 0.05 # in Amps
|
||||
ENERGY_CHANGE_SLEEP = 0.0 # put the Enerrgy settling time if needed
|
||||
FIELD_CHANGE_SLEEP = 0.0 # put the Field settling time if needed
|
||||
ACC_TIME = 1.0 # accumulation time
|
||||
|
||||
# List of scans. Each scan is defined as: (start,stop, step)
|
||||
# to make a whole loop two scans are needed: [(B1, B2, BSTEP),(B2,B1,-BSTEP)]
|
||||
# to make half of a loop one scan is needed: [(B1, B2, BSTEP),] <- don't forget to put comma before ]
|
||||
RANGES = [(B1, B2, -BSTEP),(B2,B1,BSTEP)]
|
||||
|
||||
|
||||
#Pre-actions
|
||||
# Here polarization and offsets are set
|
||||
|
||||
caput ("X11PHS-E:OPT","PGM+ID2")
|
||||
#caput(OTF_OFF1,-50) #detune ID1
|
||||
caput(OTF_OFF2,OFFSET2)
|
||||
wait_channel(OTF_DONE, 1, type = 'i')
|
||||
|
||||
if MODE is 1 or 2:
|
||||
caput(OTF_MODE2,MODE)
|
||||
else:
|
||||
raise Exception("Invalid polarization type: " + MODE)
|
||||
|
||||
wait_channel(OTF_DONE, 1, type = 'i')
|
||||
|
||||
|
||||
# setting number of samples to be averaged, which depends on the accumulation time
|
||||
avg = ACC_TIME * 10
|
||||
caput("X11MA-ES1-10ADC:AVG",avg)
|
||||
|
||||
# Generating a list of mag. fields (setpoints) needed for the loop
|
||||
setpoints = []
|
||||
for r in RANGES:
|
||||
setpoints = setpoints + frange(r[0], r[1], r[2], True)
|
||||
|
||||
# plot properties
|
||||
set_preference(Preference.ENABLED_PLOTS, ['field', 'tey_norm'])
|
||||
set_preference(Preference.PLOT_TYPES, {'tey_norm':1})
|
||||
|
||||
#scan = ManualScan(['field', 'Energy'], ['TEY', 'I0', 'polarization', 'temperature', 'RingCurrent', 'tey_norm'] , [min(setpoints), min(ENERGIES)], [max(setpoints), max(ENERGIES)], [len(setpoints)-1, len(ENERGIES)-1])
|
||||
scan = ManualScan(['field', 'Energy'], ['I0', 'TEY', 'TFY', 'polarization', 'temperature', 'RingCurrent', 'tey_norm','tfy_norm'] , [min(setpoints), min(ENERGIES)], [max(setpoints), max(ENERGIES)], [len(setpoints)-1, len(ENERGIES)-1])
|
||||
scan.start()
|
||||
|
||||
# Main loop
|
||||
for B in setpoints:
|
||||
print "Setting field = ", B
|
||||
caput("X11MA-XMCD:I-SETraw",B)
|
||||
time.sleep( FIELD_CHANGE_SLEEP ) # Settling time
|
||||
|
||||
#readback1 = caget("X11MA-XMCD:Ireadout")
|
||||
|
||||
#while abs(readback1-B) > FIELD_PRECISION:
|
||||
# readback1 = caget("X11MA-XMCD:Ireadout")
|
||||
# time.sleep(0.5)
|
||||
|
||||
|
||||
#edge
|
||||
print "Setting energy = ", E[0]
|
||||
caput(OTF_ESET, E[0])
|
||||
wait_channel(OTF_DONE, 1, type = 'i')
|
||||
readback2 = energy.read()
|
||||
if abs(readback2 - E[0]) > 0.1 : # TODO: Check accuracy
|
||||
raise Exception('Energy could not be set to the value ' + str(E(0)))
|
||||
sleep( ENERGY_CHANGE_SLEEP ) # Settling time
|
||||
|
||||
detector1 = keithley_1a.read() #Keithley1
|
||||
detector2 = keithley_2a.read() #Keithley2
|
||||
detector4 = caget(OTF_MODE2) #polarization in ID2
|
||||
M1 = detector2/detector1
|
||||
#converting polarization strings to integer numbers: C+ -> 1 and C- -> 2
|
||||
if detector4 == 'CIRC +': detector4 = 1
|
||||
elif detector4 == 'CIRC -': detector4 = 2
|
||||
else: detector4 = 0
|
||||
|
||||
|
||||
#preedge
|
||||
print "Setting energy = ", E[1]
|
||||
caput(OTF_ESET, E[1])
|
||||
wait_channel(OTF_DONE, 1, type = 'i')
|
||||
readback2 = energy.read()
|
||||
if abs(readback2 - E[1]) > 0.1 : # TODO: Check accuracy
|
||||
raise Exception('Energy could not be set to the value ' + str(E[0]))
|
||||
sleep( ENERGY_CHANGE_SLEEP ) # Settling time
|
||||
|
||||
detector1 = keithley_1a.read() #Keithley1
|
||||
detector2 = keithley_2a.read() #Keithley2
|
||||
detector4 = caget(OTF_MODE2) #polarization in ID2
|
||||
M2 = detector2/detector1
|
||||
#converting polarization strings to integer numbers: C+ -> 1 and C- -> 2
|
||||
if detector4 == 'CIRC +': detector4 = 1
|
||||
elif detector4 == 'CIRC -': detector4 = 2
|
||||
else: detector4 = 0
|
||||
|
||||
tey_norm = M2/M1;
|
||||
|
||||
#scan.append ([B, E], [readback1, readback2], [detector1, detector2, detector4, detector6, detector7, tey_norm])
|
||||
scan.append ([B, E], [readback1, readback2], [detector1, detector2, detector3, detector4, detector6, detector7, tey_norm, tfy_norm])
|
||||
|
||||
scan.end()
|
||||
caput("X11MA-ES1-10ADC:AVG",1)
|
||||
caput("X11MA-XMCD:I-SETraw",0)
|
||||
|
||||