wiki/Athos-Intensity-Diagnostics/SATFE10-PDIM067.md
2024-11-25 11:45:44 +01:00

2.2 KiB

Drawings

Diode

10 mm x 10 mm, Si PIN photodiode, S3590-09, specs

ADC

Ioxos 2111, DC coupled, 16 bit, 250 Msps, specs

Basic controls

Move diode in/out:

  1. In - caput SATFE10-PDIM067:PROBE_SP 1
  2. Out - caput SATFE10-PDIM067:PROBE_SP 0

Overview panels:
caqtdm S_SATFE10_PDIM067_expert.ui

Data

Bias: Bias is from a 9V battery and HV supply filter to the cathode Gain paths: The signal from the anode passes a signal filter and then a tee. 1 path goes to an amplifier and then a second tee. 1 branch goes to channel 0 the second brann via 20 dB attenuator to channel 2. the other branch hAS 10 db attenuator.

The following PVs are the integrated signal waveforms.

  • SATFE10-LSCP2-FNS:CH1:VAL_GET
  • SATFE10-LSCP2-FNS:CH2:VAL_GET
  • SATFE10-LSCP2-FNS:CH3:VAL_GET

Alignment

Alignment check 8.1.20
In position:

  • 40.0 mm with User offset at 0.0

Out position:

  • 0.0 mm with User offset at 0.0

Triggering

EVR:

  • SATFE10-PDIM067-EVR0

Triggering:

  • Pulser 2
  • Event 52
  • Delay 1002.612 uS
  • Width 10.000 uS
  • FrontUnivOut 6

Control:

  • On caput SATFE10-PDIM067-EVR0:FrontUnivOut6-Ena-SP 0
  • Off caput SATFE10-PDIM067-EVR0:FrontUnivOut6-Ena-SP 1