wiki/Athos-spatial-diagnostics/SATOP31-PSRD175.md
2024-12-04 15:05:16 +01:00

2.3 KiB

Drawings

Detector - Micro channel plate:

Double stack, F2224-21P159 32 mm effective area

Imaging

High voltage set points

  • MCP_front set point -1500 V
  • Screen set point 3000 V

High voltage set control

Voltage ramp time from turn on/off approx 2 minutes

Turn on
MCP front: caput SATOP31-PSRD175:CH0-ONOFF-SP 1
Screen: caput SATOP31-PSRD175:CH1-ONOFF-SP 1
Turn off
MCP front: caput SATOP31-PSRD175:CH0-ONOFF-SP 0
Screen: caput SATOP31-PSRD175:CH1-ONOFF-SP 0

Config

  • Image preview PV : SATOP31-PSRD175:FPICTURE
  • Trigger currently from SGE-CPCW-B1-EVR0, FrontUnivOut11

Stage positions

  • In: 10 mm
  • Out:70 mm

Triggering

03.11.23

  • Event: 52
  • Pulser: 6
  • Total delay: 6499.006 + 1003.011 = 7502.017 us
  • Width: 5.000 us

BS setup

4.12.24 Sources list updated
{"stream":"tcp://daqsf-daqsync-17.psi.ch:9003","split":6,"backend":"sf-imagebuffer", "labels": ["SATOP31-PSRD175"], "groups": ["Photonics"]}