wiki/Aramis-spatial-diagnostics/SAROP21-PSRD103.md

2.5 KiB

Drawings

Detector - Micro channel plate:

Double stack, F2221 - 27P320

Imaging

High voltage set points

  • MCP_front set point -1500 V
  • Screen set point 3000 V

High voltage set control

Voltage ramp time from turn on/off approx 2 minutes

Turn on
MCP front: caput SAROP21-PSRD103:CH0-ONOFF-SP 1
Screen: caput SAROP21-PSRD103:CH1-ONOFF-SP 1
Turn off
MCP front: caput SAROP21-PSRD103:CH0-ONOFF-SP 0
Screen: caput SAROP21-PSRD103:CH1-ONOFF-SP 0

Config

  • Image preview PV : SAROP21-PSRD103:FPICTURE
  • Trigger currently from SGE-CPCW-74-EVR0, FrontUnivOut9

Camera view

The camera does not image the full MCP screen (aprrox 70 %). Below are screen shots explaining what is seen by the camera.

Triggering

29.06.20

  • Event: 13
  • Pulser: 0
  • Total delay: 6498.999 + 500 = 6998.999 us
  • Width: 5.000 us