ANODE names its ranked peaks after the SFAC element it was given, so the labels read S1, S2, ... on a sulfur case but FE1, MN1, ZN1, SE1 as soon as the model carries a heavier scatterer. The parser matched "S" followed by digits, so on any such dataset it dropped the ENTIRE peak list and the dataset failed the gate as "no peaks" - however strong its signal actually was. One heme case reported no peaks when its true top peak is 13.79 sigma at 3.02x the off-site floor. The bug is silent and it mis-gates exactly the datasets most likely to widen the arbiter set, since a heavy scatterer is what makes a weakly diffracting crystal usable as an arbiter in the first place. Byte-identical on the standing all-sulfur set, verified by diffing the gate table before and after. Co-Authored-By: Claude Opus 5 (1M context) <noreply@anthropic.com>