# Detector geometry
|
|
|
|
At the moment Jungfraujoch supports solely flat detectors. The default option is to place modules in their actual location
|
|
vs. detector frame. It is not recommended to place detector modules stacked.
|
|
|
|
The simplest case is detector perpendicular to the beam. In this case it is enough to provide beam center, detector distance
|
|
and wavelength.
|
|
|
|
For more complex case, one can provide tilt of the detector rotation in PyFAI convention.
|
|
This convention uses Point Of Nominal Interaction (PONI) definition. Beam X and Y would correspond to the location on the detector,
|
|
where beam from the sample is perpendicular to the detector surface and not to the actual direct beam location. Then tilt of the detector
|
|
is defined with three rotation angles: `rot1` (rotating detector right), `rot2` (rotating detector downwards), `rot3` (rotating detector clockwise).
|
|
See [PyFAI documentation](https://pyfai.readthedocs.io/en/stable/) for more details.
|
|
|
|
## Macromolecular crystallography convention for the vertical direction
|
|
One place of confusion is the convention to have point (0,0) of the detector in the top left corner of the detector,
|
|
with Y values increasing downwards. This is also consistent with computer image formats.
|
|
|
|
However, other techniques (as well as internal operation of PSI X-ray detectors) might follow convention, for point (0,0)
|
|
being in the bottom left corner and Y values increasing upwards. Such a convention is used, for example, by PyFAI.
|
|
|
|
In general, convention is controlled in Jungfraujoch with a setting in the JSON configuration file, which allows mirroring detector in Y.
|
|
|
|
Extra care has to be taken by the user to ensure that no errors are made. |