3d4209d80365ac62632eda4a466cde7addf62ae5
Changelog entries for the three changes above, and a new CPU_DATA_ANALYSIS section on determining detector geometry from powder rings: why a ring is an independent constraint on the beam centre (it has no crystal orientation to be gauge-coupled to, unlike everything else that fits geometry here), what a ring can and cannot determine, and how the ring points are obtained. The section states the harmonics correctly, which is worth writing down because the intuitive version is wrong: a detector tilt shows up as cos(phi), the same harmonic as a beam-centre error, and the two are separated by the amplitude growing as the ring radius SQUARED - so it takes at least two rings, and on one ring they are exactly degenerate. The genuine cos(2 phi) term is hundredths of a pixel. Co-Authored-By: Claude Opus 5 (1M context) <noreply@anthropic.com>
Jungfraujoch
Application to receive data from the PSI JUNGFRAU and EIGER detectors.
All documentation is now placed in docs/ subdirectory and for the current version hosted on Jungfraujoch Read The Docs page.