The rotation geometry post-refine committed the detector distance and cell scale on any cross-validated improvement within a loose 3 % / 2 % bound. On a multi-lattice crystal the second lattice's spots bias every cross-validation fold the same way, so the "it improved" test cannot catch it and a spurious distance<->cell pair (their radial degeneracy) is committed far off the true geometry, losing reflections in the re-integration. Tighten the commit bound to < 1 % (a calibrated header needs < ~0.6 %). The move SIZE discriminates a genuine header correction from the multi-lattice failure far better than the fit residual, which genuinely marginal (noisy or ice-ring) data shares with the multi-lattice case - so a real ~0.3 % correction is kept while the ~1.7 % multi-lattice one is rejected. Co-Authored-By: Claude Opus 4.8 (1M context) <noreply@anthropic.com>