The voltage range for the repaired DENEX 250TN detector is now +2800 and -1000.
This commit is contained in:
@@ -6,15 +6,18 @@ if {$::sim_mode == "true"} {
|
||||
EvFactory new dhv1 sim
|
||||
} else {
|
||||
# clientput "Detector Voltage control not yet available"
|
||||
# MJL 6/13 Revised voltage settings for serviced original Platypus detector (rectangular DENEX 250TN)
|
||||
# Old voltages +3350V and -1500V, after servicing gas pressure changed and now +2800V and -1000V.
|
||||
# NOTE: If square DENEX 300TN is fitted, voltages need to be revised again! (+2850V and -1100V currently for this detector)
|
||||
makeasyncqueue acq NHQ200 ca2-platypus 4001
|
||||
evfactory new dhv1 nhq200 acq 1
|
||||
dhv1 tolerance 5
|
||||
dhv1 rate 25
|
||||
dhv1 upper 3350
|
||||
dhv1 upper 2800
|
||||
#dhv1 lock
|
||||
evfactory new dhv2 nhq200 acq 2
|
||||
dhv2 tolerance 5
|
||||
dhv2 rate 25
|
||||
dhv2 upper 1500
|
||||
dhv2 upper 1000
|
||||
#dhv2 lock
|
||||
}
|
||||
|
||||