From 923f056b2c4044ec259b644dfa6e24d4a7674239 Mon Sep 17 00:00:00 2001 From: Ferdi Franceschini Date: Mon, 17 Jun 2013 08:28:45 +1000 Subject: [PATCH] The voltage range for the repaired DENEX 250TN detector is now +2800 and -1000. --- .../instrument/reflectometer/config/hmm/detector.tcl | 7 +++++-- 1 file changed, 5 insertions(+), 2 deletions(-) diff --git a/site_ansto/instrument/reflectometer/config/hmm/detector.tcl b/site_ansto/instrument/reflectometer/config/hmm/detector.tcl index 3c90211e..dfaac749 100644 --- a/site_ansto/instrument/reflectometer/config/hmm/detector.tcl +++ b/site_ansto/instrument/reflectometer/config/hmm/detector.tcl @@ -6,15 +6,18 @@ if {$::sim_mode == "true"} { EvFactory new dhv1 sim } else { # clientput "Detector Voltage control not yet available" +# MJL 6/13 Revised voltage settings for serviced original Platypus detector (rectangular DENEX 250TN) +# Old voltages +3350V and -1500V, after servicing gas pressure changed and now +2800V and -1000V. +# NOTE: If square DENEX 300TN is fitted, voltages need to be revised again! (+2850V and -1100V currently for this detector) makeasyncqueue acq NHQ200 ca2-platypus 4001 evfactory new dhv1 nhq200 acq 1 dhv1 tolerance 5 dhv1 rate 25 -dhv1 upper 3350 +dhv1 upper 2800 #dhv1 lock evfactory new dhv2 nhq200 acq 2 dhv2 tolerance 5 dhv2 rate 25 -dhv2 upper 1500 +dhv2 upper 1000 #dhv2 lock }