+ make 'limit' usable as type in cfg files + minor fixes Change-Id: Ib94b2645c7a0d978d64d4c86c4415d4b5b0d485f Reviewed-on: https://forge.frm2.tum.de/review/c/sine2020/secop/playground/+/21485 Tested-by: Enrico Faulhaber <enrico.faulhaber@frm2.tum.de> Tested-by: JenkinsCodeReview <bjoern_pedersen@frm2.tum.de> Reviewed-by: Enrico Faulhaber <enrico.faulhaber@frm2.tum.de>
94 lines
2.5 KiB
INI
94 lines
2.5 KiB
INI
[node cci3he1]
|
|
description = [sim] cci3he box of MLZ Sample environment group
|
|
.
|
|
Controls an 3He insert with an ls370 and an PLC controlling
|
|
the compressor and the valves of the gas handling.
|
|
|
|
|
|
[interface tcp]
|
|
type=tcp
|
|
bindto=0.0.0.0
|
|
bindport=10767
|
|
|
|
[module T_cci3he1]
|
|
class=secop.simulation.SimDrivable
|
|
.description=Main temperature control node of cci3he1.
|
|
.
|
|
Controls the regulation loop of the ls370.
|
|
value.datatype={"type":"double","unit":"K"}
|
|
value.default=300
|
|
target.datatype={"type":"double", "min":0, "max":300, "unit":"K"}
|
|
target.default=300
|
|
.extra_params=ramp
|
|
ramp.datatype={"type":"double","min":0,"max":600,"unit":"K/min"}
|
|
ramp.description=target ramping speed in K/min.
|
|
ramp.default=60
|
|
.meaning=["temperature_regulation",40]
|
|
|
|
[module T_cci3he1_A]
|
|
class=secop.simulation.SimReadable
|
|
.description=3He pot temperature sensor. Also used for the regulation.
|
|
.visibility=expert
|
|
value.default=300
|
|
value.datatype={"type":"double","unit":"K"}
|
|
.meaning=["temperature",38]
|
|
|
|
[module T_cci3he1_B]
|
|
class=secop.simulation.SimReadable
|
|
.description=(optional) sample temperature sensor close to sample.
|
|
.visibility=user
|
|
value.default=300
|
|
value.datatype={"type":"double","unit":"K"}
|
|
.meaning=["temperature",39]
|
|
|
|
[module cci3he1_p1]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure at turbo pump inlet.
|
|
.visibility=expert
|
|
value.default=2e-3
|
|
value.datatype={"type":"double","unit":"mbar"}
|
|
|
|
[module cci3he1_p2]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure at turbo pump outlet.
|
|
.visibility=expert
|
|
value.default=9.87
|
|
value.datatype={"type":"double","unit":"mbar"}
|
|
|
|
[module cci3he1_p3]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure at compressor inlet.
|
|
.visibility=expert
|
|
value.default=19.99
|
|
value.datatype={"type":"double","unit":"mbar"}
|
|
|
|
[module cci3he1_p4]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure at compressor outlet.
|
|
.visibility=expert
|
|
value.default=999
|
|
value.datatype={"type":"double","unit":"mbar"}
|
|
|
|
[module cci3he1_p5]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure in dump tank.
|
|
.visibility=expert
|
|
value.default=567
|
|
value.datatype={"type":"double","unit":"mbar"}
|
|
|
|
[module cci3he1_p6]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure in the vacuum dewar (ivc).
|
|
.visibility=expert
|
|
value.default=1e-3
|
|
value.datatype={"type":"double","unit":"mbar"}
|
|
|
|
[module cci3he1_flow]
|
|
class=secop.simulation.SimReadable
|
|
.description=Gas Flow (condensing line).
|
|
.visibility=expert
|
|
value.default=12.34
|
|
value.datatype={"type":"double","unit":"mln/min"}
|
|
|
|
# note: all valves and switches are missing: use VNC to control them
|