+ make 'limit' usable as type in cfg files + minor fixes Change-Id: Ib94b2645c7a0d978d64d4c86c4415d4b5b0d485f Reviewed-on: https://forge.frm2.tum.de/review/c/sine2020/secop/playground/+/21485 Tested-by: Enrico Faulhaber <enrico.faulhaber@frm2.tum.de> Tested-by: JenkinsCodeReview <bjoern_pedersen@frm2.tum.de> Reviewed-by: Enrico Faulhaber <enrico.faulhaber@frm2.tum.de>
108 lines
3.0 KiB
INI
108 lines
3.0 KiB
INI
[node ccidu1]
|
|
description = [sim] ccidu box of MLZ Sample environment group
|
|
.
|
|
Controls an 3He/4He dilution insert with an ls372 and an PLC controlling
|
|
the compressor and the valves of the gas handling.
|
|
|
|
|
|
[interface tcp]
|
|
type=tcp
|
|
bindto=0.0.0.0
|
|
bindport=10767
|
|
|
|
[module T_ccidu1]
|
|
class=secop.simulation.SimDrivable
|
|
.description=Main temperature control node of ccidu1.
|
|
.
|
|
Controls the regulation loop of the ls372.
|
|
value.unit='K'
|
|
value.default=300
|
|
target.datatype={"type":"double", "min":0, "max":300, "unit":"K"}
|
|
target.default=300
|
|
.extra_params=ramp
|
|
ramp.datatype={"type":"double", "min":0, "max":600, "unit":"K/min"}
|
|
ramp.description=target ramping speed in K/min.
|
|
ramp.default=60
|
|
.meaning=["temperature_regulation",40]
|
|
|
|
[module T_ccidu1_A]
|
|
class=secop.simulation.SimReadable
|
|
.description=mixing chamber temperature sensor. Also used for the regulation.
|
|
.visibility=expert
|
|
value.default=300
|
|
value.datatype={"type":"double", "unit":"K"}
|
|
.meaning=["temperature",38]
|
|
|
|
[module T_ccidu1_B]
|
|
class=secop.simulation.SimReadable
|
|
.description=(optional) sample temperature sensor close to sample.
|
|
.visibility=user
|
|
value.default=300
|
|
value.datatype={"type":"double", "unit":"K"}
|
|
.meaning=["temperature",39]
|
|
|
|
[module ccidu1_pstill]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure at the still/turbo pump inlet.
|
|
.visibility=expert
|
|
value.default=999
|
|
value.datatype={"type":"double", "unit":"mbar"}
|
|
|
|
[module ccidu1_pinlet]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure at forepump inlet/turbo pump outlet.
|
|
.visibility=expert
|
|
value.default=999
|
|
value.datatype={"type":"double", "unit":"mbar"}
|
|
|
|
[module ccidu1_poutlet]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure at forepump outlet/compressor inlet.
|
|
.visibility=expert
|
|
value.default=999
|
|
value.datatype={"type":"double", "unit":"mbar"}
|
|
|
|
[module ccidu1_pkond]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure at condensing line/compressor outlet.
|
|
.visibility=expert
|
|
value.default=999
|
|
value.datatype={"type":"double", "unit":"mbar"}
|
|
|
|
[module ccidu1_ptank]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure in dump tank.
|
|
.visibility=expert
|
|
value.default=999
|
|
value.datatype={"type":"double", "unit":"mbar"}
|
|
|
|
[module ccidu1_pvac]
|
|
class=secop.simulation.SimReadable
|
|
.description=Pressure in the vacuum dewar (ivc).
|
|
.visibility=expert
|
|
value.default=999
|
|
value.datatype={"type":"double", "unit":"mbar"}
|
|
|
|
[module ccidu1_flow]
|
|
class=secop.simulation.SimReadable
|
|
.description=Gas Flow (condensing line).
|
|
.visibility=expert
|
|
value.default=999
|
|
value.datatype={"type":"double", "unit":"mbar"}
|
|
|
|
# note: all valves and switches are missing: use VNC to control them
|
|
[module ccidu1_V6]
|
|
class=secop.simulation.SimDrivable
|
|
.description=Needle valve
|
|
.visibility=expert
|
|
value.default=99
|
|
value.datatype={"type":"double", "min":0, "max":100, "unit":"%%"}
|
|
|
|
[module ccidu1_V3]
|
|
class=secop.simulation.SimWritable
|
|
.description=Dump Valve
|
|
.visibility=expert
|
|
value.default="OFF"
|
|
value.datatype={"type":"enum", "members":{"on": 1, "OFF":0}}
|
|
target.datatype={"type":"enum", "members":{"on": 1, "OFF":0}}
|