rework switch timing
- specific things in ips_mercury.py - general things in magfield.py Change-Id: I7c2bae815b9a80a17803b44b8941ef3dea3adb60
This commit is contained in:
parent
c3f55435da
commit
b0315e133b
@ -20,6 +20,7 @@
|
||||
# *****************************************************************************
|
||||
"""oxford instruments mercury IPS power supply"""
|
||||
|
||||
import time
|
||||
from secop.core import Parameter, EnumType, FloatRange, BoolType
|
||||
from secop.lib.enum import Enum
|
||||
from secop.errors import BadValueError, HardwareError
|
||||
@ -51,6 +52,7 @@ class Field(MercuryChannel, Magfield):
|
||||
nslaves = 3
|
||||
slave_currents = None
|
||||
__init = True
|
||||
__reset_switch_time = False
|
||||
|
||||
def doPoll(self):
|
||||
super().doPoll()
|
||||
@ -62,7 +64,7 @@ class Field(MercuryChannel, Magfield):
|
||||
if self.__init:
|
||||
self.__init = False
|
||||
self.persistent_field = pf
|
||||
if self.switch_heater != 0 or self._field_mismatch is None:
|
||||
if self.switch_heater == self.switch_heater.on or self._field_mismatch is None:
|
||||
self.forced_persistent_field = False
|
||||
self._field_mismatch = False
|
||||
return self.current
|
||||
@ -94,7 +96,18 @@ class Field(MercuryChannel, Magfield):
|
||||
return self.change('PSU:ACTN', value, hold_rtoz_rtos_clmp)
|
||||
|
||||
def read_switch_heater(self):
|
||||
return self.query('PSU:SIG:SWHT', off_on)
|
||||
value = self.query('PSU:SIG:SWHT', off_on)
|
||||
now = time.time()
|
||||
switch_time = self.switch_time[self.switch_heater]
|
||||
if value != self.switch_heater:
|
||||
self.__reset_switch_time = True
|
||||
if now < (switch_time or 0) + 10:
|
||||
# probably switch heater was changed, but IPS reply is not yet updated
|
||||
return self.switch_heater
|
||||
elif self.__reset_switch_time:
|
||||
self.__reset_switch_time = False
|
||||
self.switch_time = [None, None]
|
||||
return value
|
||||
|
||||
def write_switch_heater(self, value):
|
||||
return self.change('PSU:SIG:SWHT', value, off_on)
|
||||
|
@ -44,6 +44,9 @@ Status = Enum(
|
||||
FINALIZING=390,
|
||||
)
|
||||
|
||||
OFF = 0
|
||||
ON = 1
|
||||
|
||||
|
||||
class Magfield(HasLimits, Drivable):
|
||||
value = Parameter('magnetic field', datatype=FloatRange(unit='T'))
|
||||
@ -52,7 +55,7 @@ class Magfield(HasLimits, Drivable):
|
||||
'persistent mode', EnumType(Mode), readonly=False, default=Mode.PERSISTENT)
|
||||
tolerance = Parameter(
|
||||
'tolerance', FloatRange(0, unit='$'), readonly=False, default=0.0002)
|
||||
switch_heater = Parameter('switch heater', EnumType(off=0, on=1),
|
||||
switch_heater = Parameter('switch heater', EnumType(off=OFF, on=ON),
|
||||
readonly=False, default=0)
|
||||
persistent_field = Parameter(
|
||||
'persistent field', FloatRange(unit='$'), readonly=False)
|
||||
@ -87,8 +90,7 @@ class Magfield(HasLimits, Drivable):
|
||||
_state = None
|
||||
__init = True
|
||||
_last_target = None
|
||||
switch_on_time = None
|
||||
switch_off_time = None
|
||||
switch_time = None, None
|
||||
|
||||
def doPoll(self):
|
||||
if self.__init:
|
||||
@ -185,14 +187,11 @@ class Magfield(HasLimits, Drivable):
|
||||
|
||||
def update_switch_heater(self, value):
|
||||
"""is called whenever switch heater was changed"""
|
||||
if value != 0:
|
||||
self.switch_off_time = None
|
||||
if self.switch_on_time is None:
|
||||
self.switch_on_time = time.time()
|
||||
else:
|
||||
self.switch_on_time = None
|
||||
if self.switch_off_time is None:
|
||||
self.switch_off_time = time.time()
|
||||
switch_time = self.switch_time[value]
|
||||
if switch_time is None:
|
||||
switch_time = time.time()
|
||||
self.switch_time = [None, None]
|
||||
self.switch_time[value] = switch_time
|
||||
|
||||
def start_switch_on(self, state):
|
||||
"""switch heater on"""
|
||||
@ -213,12 +212,10 @@ class Magfield(HasLimits, Drivable):
|
||||
abs(self.target - self.persistent_field) <= self.tolerance): # short cut
|
||||
return self.check_switch_off
|
||||
self.read_switch_heater()
|
||||
if self.switch_on_time is None:
|
||||
if state.now - self.switch_off_time > 10:
|
||||
self.log.warning('switch turned off manually?')
|
||||
return self.start_switch_on
|
||||
return Retry()
|
||||
if state.now - self.switch_on_time < self.wait_switch_on:
|
||||
if self.switch_time[ON] is None:
|
||||
self.log.warning('switch turned off manually?')
|
||||
return self.start_switch_on
|
||||
if state.now - self.switch_time[ON] < self.wait_switch_on:
|
||||
return Retry()
|
||||
self._last_target = self.target
|
||||
return self.start_ramp_to_target
|
||||
@ -279,12 +276,10 @@ class Magfield(HasLimits, Drivable):
|
||||
return self.start_switch_on
|
||||
self.persistent_field = self.value
|
||||
self.read_switch_heater()
|
||||
if self.switch_off_time is None:
|
||||
if state.now - self.switch_on_time > 10:
|
||||
self.log.warning('switch turned on manually?')
|
||||
return self.start_switch_off
|
||||
return Retry()
|
||||
if state.now - self.switch_off_time < self.wait_switch_off:
|
||||
if self.switch_time[OFF] is None:
|
||||
self.log.warning('switch turned on manually?')
|
||||
return self.start_switch_off
|
||||
if state.now - self.switch_time[OFF] < self.wait_switch_off:
|
||||
return Retry()
|
||||
if abs(self.value) > self.persistent_limit:
|
||||
self.status = Status.IDLE, 'leads current at field, switch off'
|
||||
|
Loading…
x
Reference in New Issue
Block a user