Real-HW-adjacent (sim) testing surfaced three issues with the smear calibration aid: 1. The composite reverted to the plain rotating live view right before the click prompt. find_rotation_center_smear_experimental() called update_frame() after _smear_sweep(), which grabs a fresh raw live frame -- overwriting the composite it just built. Also, _push_smear_composite always resumed live mode, so even without that bug the composite would only stay visible for hold_display_s. Fixed: _push_smear_composite gained resume_live=False for the final push; _smear_sweep now takes keep_shutter_open and closes the shutter itself at the end, so the composite (not a fresh grab) stays frozen on screen for the click. The stray update_frame() call is removed. 2. The composite appeared rotated relative to the live view. get_last_image() frames already have num_rotation_90/transpose applied (PreviewSignal.put() applies it to whatever it stores), so composites built from those frames got the transform applied AGAIN by push_preview_image()'s own self.image.put() call. Fixed: push_preview_image() now undoes the transform first (reverse order, since transpose is applied last forward) so the net effect is a single transform, matching a live frame. 3. Feature request: the sim camera's rotation-coupled pattern was a single eccentric gaussian blob -- a better analog for the isolated/ point-particle case than the extended/textured case this feature actually targets. Added generate_speckle_grains/ make_speckle_test_pattern (sim_cameras.py): a fixed, seeded field of small gaussian "grains" at a range of radii, rotating rigidly together as the live angle changes -- producing several concentric arcs in the composite instead of one, much closer to a real textured sample. New sim_speckle_count/sim_speckle_seed config, defaults requiring no yaml changes. Co-Authored-By: Claude Sonnet 5 <noreply@anthropic.com>
cSAXS BEC
cSAXS specific plugins and configs for BEC This document guides you through the procedure to spin up BEC at the beamline for a new experiment (e-account). You might want to run cSAXS copy scripts before in case you want to have the former data structure to be preserved.
Overview
- Clone cSAXS BEC repository into e-account (e.g. into ~/Data10/software/.)
- Start Epics iocs
- Start BEC, BEC server and load/modify the device config with relevant hardware
- BEC commands
Clone cSAXS BEC repository
Clone the current cSAXS BEC repository from GIT into the new e-account. Create directory
mkdir ~/Data10/software
cd ~/Data10/software
Clone repository
git clone https://gitlab.psi.ch/bec/csaxs_bec.git
Start epics iocs
You can start up the iocs while the ./setup_bec.sh script is running. Be aware though that the scripts requires you to interact with it.
DelayGenerator
Open a new tab in a terminal in comp1/comp2 or cons1 and follow the commands below:
ssh gac-x12sa@localhost
cd ~/delaygen
iocsh -7.0.6 startup.script
Be aware -7.0.6 is referring to the current epics version and might change in future (SLS 2.0). To start the epics panel (only if needed), run in a new shell
caqtdm -noMsg -macro P=delaygen:,R=DG1: srsDG645.ui
More notes on usages and cabelling of DDGs. Currently 3 generators are used:
- DDG1 for detectors (EXT Enable from SGalil stages, T0 to DDG2, AB for eiger, CD for Falcon, EF for Pilatus_2)
- DDG2 for mcs card (ext. enable from DDG1, AB to mcs card)
- DDG3 for fast shutter and mcs enable (AB short pulse to enable MCS (needed!!), CD to keep FSH open )
Eiger
Open a new tab in a terminal in comp1/comp2 or cons1 and follow the commands below:
ssh gac-x12sa@localhost
cd ~/Software/Eiger/
./launch_ioc
A live view of status and images for std_daq can be found here xbl-daq-29:5000.
Pilatus_2
First, start the cam server. Connect to the computer and follow the instructions printed after connecting:
ssh det@x12sa-pd-2 (Pilatus2)
The cam server will open, wait until you see **. Nex step, open a new tab in a terminal in comp1/comp2 or cons1 and follow the commands below:
ssh gac-x12sa@localhost
cd ~/Software/Pilatus300K/
./launch_epics
FalconX1
Open a new tab in a terminal in comp1/comp2 or cons1 (logged in as the current e-account) and follow the commands:
ssh x12sa-cons-01
cd /ioc/X12SA-PC-FALCONX1/
iocsh -7.0.6 startup.script
Be aware -7.0.6 is referring to the current epics version and might change in future (SLS 2.0)
Start BEC, BEC server and load device config
Step 1 needs to have finished for continuing with these steps. What remains now is to start the bec server. Connect to pc15543 and open a new terminal to run:
cd ~/Data10/software
source bec_venv/bin/activate
bec-server start
tmux attach -t bec
Open another teminal on pc15543 and start BEC:
cd ~/Data10/software
source bec_venv/bin/activate
bec
Note: In case there is a warning after starting BEC that it was not able to import scilog, you will have to pip install scilog in the bec_venv to be able to send printouts to scilog from the command line. Within a new terminal:
cd ~/Data10/software
source bec_venv/bin/activate
pip install scilog
Device config from csaxs-bec plugins
bec.config.update_session_with_file('/sls/X12SA/data/e20639/Data10/software/csaxs-bec/bec_plugins/configs/bec_device_config_sastt.yaml')
bec.config.save_current_session('~/Data10/software/current_config.yaml')
The second command is helpful if you adjust limits of motors, which will then be stored in the config and loaded if a reload of the configuration is needed.
BEC commands
A number of commands that are useful:
To move devices that are added in the config:
umv(dev.samx, 0) #absolute
umvr(dev.samx, 0) #relative
dev.samx.wm #print motor position and limits
dev.samx.limits = [low_limit, high_limit] # To set limits, note, same lower and higher limit, e.g. [0, 0] means no limits!!
Scans:
scans.acquire(exp_time = 0.5, frames_per_trigger=10, readout_time=3e-3) # equivalent to a loopscan 30 0.5
scans.line_scan(dev.samx, -1, 1, steps=20, exp_time=0.5, readout_time=3e-3, relative=True) # optional, add frames_per_trigger =10 for burst acquisition at each step
#Scan 2 motors in a step scan at the same time
scans.line_scan(dev.samx, -1, 1, dev.samy, -1, 1, steps=20, exp_time=0.5, readout_time =3e-3, relative=True) # step scan with 20 steps, again frames_per_trigger can be added for burst at each point
scans.sgalil_grid(start_y = , end_y = , interval_y = , start_x=, end_x=, interval_x =, exp_time=0.5, readout_time=3e-3, relative=True)