x01dcandClaude Sonnet 5 39e81130af fix(LamNI): reset stale live-view channel, add GUI status/toggle switches
A Ctrl-C interrupting an in-flight RPC call (e.g. mid
set_live_view_signal) could leave the XRayEye widget stuck showing the
smear-preview channel afterward -- and since lamnigui_show_xeyealign()
reuses the existing xeyegui widget across calls rather than recreating
it, that stale state would silently carry into the *next* alignment
run too (including production align()/find_rotation_center(), since
on_live_view_enabled() now respects whichever channel is selected).

Fix: lamnigui_show_xeyealign() unconditionally resets both
set_live_view_signal("image") and set_smear_active(False) every time
it's called -- both are cheap no-ops if already in the default state,
and this is the single shared entry point for every alignment
procedure, so one fix covers all of them.

Also add two GUI-facing controls to XRayEye, in a new row below the
shutter/camera-running toggles:
- "Smear preview channel" -- an interactive toggle mirroring
  set_live_view_signal(), so the operator has a GUI-only recovery path
  independent of the ipython client (e.g. if it's hung/disconnected).
- "Smear integrating" -- a read-only status indicator (disabled
  ToggleSwitch) reflecting whether _smear_sweep() is currently
  accumulating a composite, driven entirely by the automation.

set_live_view_signal() keeps the interactive toggle's visual state in
sync regardless of who changed the channel (automation or operator) --
single source of truth. _smear_sweep() brackets accumulation with
set_smear_active(True)/(False), turning it off as soon as the sweep
ends (independent of the channel switch, which stays on the composite
until the operator's click is collected). Both new XRayEye methods
required regenerating csaxs_bec/bec_widgets/widgets/client.py via
bw-generate-cli --target csaxs_bec.

Co-Authored-By: Claude Sonnet 5 <noreply@anthropic.com>
2026-07-24 16:08:01 +02:00
2025-05-24 11:37:51 +02:00
2026-07-12 23:16:18 +02:00

cSAXS BEC

cSAXS specific plugins and configs for BEC This document guides you through the procedure to spin up BEC at the beamline for a new experiment (e-account). You might want to run cSAXS copy scripts before in case you want to have the former data structure to be preserved.

Overview

  • Clone cSAXS BEC repository into e-account (e.g. into ~/Data10/software/.)
  • Start Epics iocs
  • Start BEC, BEC server and load/modify the device config with relevant hardware
  • BEC commands

Clone cSAXS BEC repository

Clone the current cSAXS BEC repository from GIT into the new e-account. Create directory

mkdir ~/Data10/software
cd ~/Data10/software

Clone repository

git clone https://gitlab.psi.ch/bec/csaxs_bec.git

Start epics iocs

You can start up the iocs while the ./setup_bec.sh script is running. Be aware though that the scripts requires you to interact with it.

DelayGenerator

Open a new tab in a terminal in comp1/comp2 or cons1 and follow the commands below:

ssh gac-x12sa@localhost
cd ~/delaygen
iocsh -7.0.6 startup.script

Be aware -7.0.6 is referring to the current epics version and might change in future (SLS 2.0). To start the epics panel (only if needed), run in a new shell

caqtdm -noMsg -macro P=delaygen:,R=DG1: srsDG645.ui

More notes on usages and cabelling of DDGs. Currently 3 generators are used:

  • DDG1 for detectors (EXT Enable from SGalil stages, T0 to DDG2, AB for eiger, CD for Falcon, EF for Pilatus_2)
  • DDG2 for mcs card (ext. enable from DDG1, AB to mcs card)
  • DDG3 for fast shutter and mcs enable (AB short pulse to enable MCS (needed!!), CD to keep FSH open )

Eiger

Open a new tab in a terminal in comp1/comp2 or cons1 and follow the commands below:

ssh gac-x12sa@localhost
cd ~/Software/Eiger/
./launch_ioc

A live view of status and images for std_daq can be found here xbl-daq-29:5000.

Pilatus_2

First, start the cam server. Connect to the computer and follow the instructions printed after connecting:

ssh det@x12sa-pd-2 (Pilatus2)

The cam server will open, wait until you see **. Nex step, open a new tab in a terminal in comp1/comp2 or cons1 and follow the commands below:

ssh gac-x12sa@localhost
cd ~/Software/Pilatus300K/
./launch_epics

FalconX1

Open a new tab in a terminal in comp1/comp2 or cons1 (logged in as the current e-account) and follow the commands:

ssh x12sa-cons-01
cd /ioc/X12SA-PC-FALCONX1/
iocsh -7.0.6 startup.script

Be aware -7.0.6 is referring to the current epics version and might change in future (SLS 2.0)

Start BEC, BEC server and load device config

Step 1 needs to have finished for continuing with these steps. What remains now is to start the bec server. Connect to pc15543 and open a new terminal to run:

cd ~/Data10/software
source bec_venv/bin/activate
bec-server start
tmux attach -t bec

Open another teminal on pc15543 and start BEC:

cd ~/Data10/software
source bec_venv/bin/activate
bec

Note: In case there is a warning after starting BEC that it was not able to import scilog, you will have to pip install scilog in the bec_venv to be able to send printouts to scilog from the command line. Within a new terminal:

cd ~/Data10/software
source bec_venv/bin/activate
pip install scilog

Device config from csaxs-bec plugins

bec.config.update_session_with_file('/sls/X12SA/data/e20639/Data10/software/csaxs-bec/bec_plugins/configs/bec_device_config_sastt.yaml')
bec.config.save_current_session('~/Data10/software/current_config.yaml')

The second command is helpful if you adjust limits of motors, which will then be stored in the config and loaded if a reload of the configuration is needed.

BEC commands

A number of commands that are useful:

To move devices that are added in the config:

umv(dev.samx, 0) #absolute
umvr(dev.samx, 0) #relative
dev.samx.wm #print motor position and limits
dev.samx.limits = [low_limit, high_limit] # To set limits, note, same lower and higher limit, e.g. [0, 0] means no limits!!

Scans:

scans.acquire(exp_time = 0.5, frames_per_trigger=10, readout_time=3e-3) # equivalent to a loopscan 30 0.5
scans.line_scan(dev.samx, -1, 1, steps=20, exp_time=0.5, readout_time=3e-3, relative=True) # optional, add frames_per_trigger =10 for burst acquisition at each step
#Scan 2 motors in a step scan at the same time
scans.line_scan(dev.samx, -1, 1, dev.samy, -1, 1, steps=20, exp_time=0.5, readout_time =3e-3, relative=True) # step scan with 20 steps, again frames_per_trigger can be added for burst at each point
scans.sgalil_grid(start_y = , end_y = , interval_y = , start_x=, end_x=, interval_x =, exp_time=0.5, readout_time=3e-3, relative=True)
S
Description
cSAXS specific plugins and configs for BEC
Readme BSD-3-Clause
24 MiB
Languages
Python 99%
HTML 1%