docs(flomni): FZP calibration / range how-tos updated for the PREEMPT_RT loop (original text kept) #337
No files matched your search
+128
@@ -0,0 +1,128 @@
|
||||
# How to calibrate the FZP scanner of flOMNI
|
||||
|
||||
Source of `local_contact_only-howto_calibrate_the_fzp_scanner.pdf`. The original text is kept unchanged; the
|
||||
section "Update 2026-10-20" at the end lists what the present flOMNI loop (PREEMPT_RT, 100 µs, branch `rt_preemt`)
|
||||
needs in addition.
|
||||
|
||||
```text
|
||||
This is a description of how to calibrate the FZP scanner of flOMNI
|
||||
|
||||
Rmemeber to do all modifications for both directions x and y!
|
||||
|
||||
1. Align flOMNI without the FZP scanner. This implies disabling the
|
||||
FZP scanner completely. To do so, in OrchestraHMI edit the parameters
|
||||
of the FZP scanner as follows.
|
||||
(a) Module PID_i_FZP: Set Param FAST_FZP_PID_LOOP(0) to 2. Then the FZP
|
||||
will just remain
|
||||
statically centered to the capacitive sensor position in closed loop.
|
||||
(b) Module PID_i: Set Param FAST_FZP_PID_LOOP(0) to 0. If this is
|
||||
negative, the FZP scanner will move ahead in a point-by-point scan by
|
||||
that amount.
|
||||
(c) Module PID_i: Set Param FAST_FZP_PID_LOOP(5) to 0. If this is
|
||||
larger than zero the sample stage motion is stopped if the position is
|
||||
reached by the FZP scanner (which is the case when the scanner is not
|
||||
active). Zero means thus full compensation by the sample stage.
|
||||
|
||||
In this state the setup can be aligned, and the affine matrix for
|
||||
ptychography can be found. This refers then purely to the alignment of
|
||||
the interferometer and X-ray detector.
|
||||
|
||||
2. In the next step we perform a scan (with a small beam and small
|
||||
range) such that only the FZP scanner is used and the sample stage
|
||||
will not move.
|
||||
|
||||
Before we start ensure that the Correction Matrix for the capacitive
|
||||
sensors is cleared.
|
||||
|
||||
To do this change the parameters of the flOMNI control
|
||||
loop as follows:
|
||||
|
||||
(a) Module AnalogInput_6259: Set parameter MatrixCorr to 1 0 0 1
|
||||
(b) Module PID_j_fzp: Set Param FAST_FZP_PID_LOOP(0) to 1 to re-activate
|
||||
the FZP scanner.
|
||||
(c) Module PID_i: Set Param FAST_FZP_PID_LOOP(5) to 10, so up to 10
|
||||
volts output of the fast PID is accepted before starting a movement of
|
||||
the sample stage.
|
||||
(d) Module PID_i: Param FAST_FZP_PID_LOOP(4) 0
|
||||
(e) Module PID_i: Param FAST_FZP_PID_LOOP(6) 1
|
||||
with d and e it will now stabilize the sample stage position at zero
|
||||
(f) Module PID_i: Param FAST_FZP_PID_LOOP(0) remains 0 (no move ahead
|
||||
in a scan)
|
||||
|
||||
With these settings perform a scan. Use a field of view matching the
|
||||
range of the FZP scanner. Watch the sample stage position in Orchestra
|
||||
to ensure it does not change!
|
||||
A good range should be 2.8 microns in x and 1.8 microns in y.
|
||||
Reconstruct this scan with Michal's position refinement code and ask
|
||||
for an update of the FZP matrix. This can then be put into the
|
||||
AnalogInput_6259 module.
|
||||
I guess one has to have the matrix found for the interferometer in the
|
||||
reconsturction code, such that the new matrix of the FZP scanner will
|
||||
adjust the coordinate system of the FZP scanner to the system of the
|
||||
interferometer. Later these positions are used together and should
|
||||
match in direction.
|
||||
|
||||
3. Back to default operating values:
|
||||
(a) Module PID_i: Set Param FAST_FZP_PID_LOOP(0) to -0.5, for up to
|
||||
0.5 microns move ahead in a scan.
|
||||
All other parameters are already as supposed to be.
|
||||
Default values for the sample stage pid, fast FZP parameter are "-
|
||||
0.5,0,0,0,0.1,0.5,0"
|
||||
Default values for the fzp pid, fast FZP parameter are "1,0.0001,0.1,-
|
||||
4,0,0,0"
|
||||
```
|
||||
|
||||
## Update 2026-10-20 (flOMNI loop on PREEMPT_RT, mpc2844)
|
||||
|
||||
The loop has new modules since the text above was written (flOMNI `FeedbackLoop/tuning_guide.md`,
|
||||
`feedback_report.md` 5.6, 6n, 6o, 6q). Parameters are changed in the web HMI (`http://mpc2844:8080`, tab
|
||||
Parameters) instead of OrchestraHMI. `i` = x or y, as above. Module names now: `PID_x`, `PID_y` (sample stage),
|
||||
`PID_x_fzp`, `PID_y_fzp` (FZP scanner), `AnalogInput_6289` (was `AnalogInput_6259`).
|
||||
|
||||
**FAST_FZP_PID_LOOP has 7 elements** (`PID.cpp`). Sample stage PIDs: (0) move-ahead (negative, µm, point-by-point
|
||||
scans only), (4) FZP error threshold (µm) and (5) FZP drive limit (V) for holding the sample stage while the FZP
|
||||
takes the step (5 = 0: never held), (6) 1 = hold the sample stage at 0. FZP PIDs: (0) mode (1 operation, 2 FZP
|
||||
centred on its cap sensor = disabled, 3 not commissioned, 4 debug), (1)–(3) min / max / gain of the old velocity
|
||||
correction (−4 before 2026-10-19, now 0: replaced by the feedforward).
|
||||
|
||||
**In addition to step 1 (FZP disabled):**
|
||||
|
||||
- `Feedforward`: `enable` 0,0 — the static FZP feedforward adds its drive also in FZP mode 2.
|
||||
- `AbbeCorrection`: `enable` 0 (default until the signs are known).
|
||||
- `TargetSelect_x`, `TargetSelect_y`: `use_final` 0; `OnTargetTrigger`: `WaitForSlew` 1 (defaults).
|
||||
- `FreqGen`: `enable` 0 (excitation for measurements; default).
|
||||
|
||||
**In addition to step 2 (FZP-only calibration scan):**
|
||||
|
||||
- FZP range: `PID_x_fzp` / `PID_y_fzp` `Ub` / `Lb` are ±1.5 V in operation (the cap sensors are calibrated over
|
||||
about ±2 V only). For the field of view above (2.8 × 1.8 µm) widen them, e.g. to ±10 V (FZP ≈ 0.13 µm/V in x,
|
||||
≈ 0.05 µm/V in y). The feedforward stays off (`enable` 0,0) during this scan.
|
||||
- The detector trigger (`OnTargetTrigger`) checks the relative error sample vs FZP (tolerance / hold time from the
|
||||
ConstEmitter, BEC `interferometer_trigger_*`, default 10 nm / 3 ms) and waits `MinDelay_ms` (15 ms).
|
||||
|
||||
**Step 3, default operating values (replace the values in the text above):**
|
||||
|
||||
| Module | Parameter | Value |
|
||||
|---|---|---|
|
||||
| `PID_x`, `PID_y` | `FAST_FZP_PID_LOOP` | `0,0,0,0,0.1,0,0` (no move-ahead: −0.5 µm does not fit the ±1.5 V FZP range) |
|
||||
| `PID_x_fzp` | `FAST_FZP_PID_LOOP`, `Ub` / `Lb` | `1,0.0001,0.1,0,0,0,0`, ±1.5 V |
|
||||
| `PID_y_fzp` | `FAST_FZP_PID_LOOP`, `Ub` / `Lb` | `1,0.0005,0.02,0,0,0,0`, ±1.5 V |
|
||||
| `Feedforward` | `enable`; `ff_limit`, `out_min` / `out_max` | 1,1; ±1.5 V (as the FZP `Ub` / `Lb`) |
|
||||
| `OnTargetTrigger` | `WaitForSlew`, `MinDelay_ms` | 1, 15 |
|
||||
| `TargetSelect_x`, `TargetSelect_y` | `use_final` | 0 |
|
||||
| `AbbeCorrection` | `enable` | 0 (until the signs are known) |
|
||||
| `AnalogInput_6289` | `MatrixCorr` | the new matrix from the calibration |
|
||||
|
||||
Check afterwards: web HMI tab **Loop check**, level 1 (it expects these values and lists live values that differ
|
||||
from the `.xmp` files); save the new values as default (Parameters → Unsaved changes) or in the `.xmp`.
|
||||
|
||||
**After the calibration (per experiment): faster scans, exposure during the motion.** With the FZP calibrated
|
||||
over a wider range the FZP can take the small steps alone (flOMNI `feedback_report.md` 6q, tested at ±7 V):
|
||||
|
||||
- `PID_x_fzp` / `PID_y_fzp` `Ub` / `Lb` and the `Feedforward` limits to the calibrated range (e.g. ±7 V);
|
||||
- `TargetSelect_x` / `_y` `use_final` 1 (the FZP aims at the new point at once, the sample stage follows its
|
||||
S-curve) and `OnTargetTrigger` `WaitForSlew` 0 (trigger on the relative error only); optionally `MinDelay_ms` 0;
|
||||
- optionally move-ahead, sized to the FZP range: `PID_x` (0) e.g. −0.5, `PID_y` (0) e.g. −0.2.
|
||||
|
||||
Measured with 101 ms exposures: 0.2–1 µm steps trigger after 4.5–5.8 ms instead of 17–32 ms; the error during the
|
||||
exposure grows with the step (1 µm: ≈ 6 nm max instead of 2.5 nm), so the setting depends on the experiment.
|
||||
BIN
Binary file not shown.
+36
@@ -0,0 +1,36 @@
|
||||
# How to limit the FZP scanner range of flOMNI
|
||||
|
||||
Source of `local_contact_only-howto_limit_the_fzp_scanner_range.pdf`. The original text is kept unchanged; the
|
||||
section "Update 2026-10-20" lists what the present flOMNI loop (PREEMPT_RT, 100 µs, branch `rt_preemt`) needs in
|
||||
addition.
|
||||
|
||||
```text
|
||||
to limit the fzp scanner ranges
|
||||
in PID_ FAST FZP PID LOOP
|
||||
0: reduce to small ahead movement range, such as -0.05
|
||||
4: reduce error threshold, eg 0.01
|
||||
5: inform PID that FZP range is small, such as 0.1
|
||||
|
||||
in PID_ _FZP
|
||||
limit the analog output LB and UB
|
||||
```
|
||||
|
||||
## Update 2026-10-20 (flOMNI loop on PREEMPT_RT, mpc2844)
|
||||
|
||||
The limited range is the present operating state: `PID_x_fzp` / `PID_y_fzp` `Ub` / `Lb` = ±1.5 V because the cap
|
||||
sensors are calibrated over about ±2 V only (flOMNI `feedback_report.md` 6n). FZP ≈ 0.13 µm/V in x, ≈ 0.05 µm/V in y,
|
||||
so ±1.5 V ≈ ±0.19 µm (x) / ±0.08 µm (y). In the sample stage PIDs (`PID_x`, `PID_y`) `FAST_FZP_PID_LOOP` is
|
||||
`0,0,0,0,0.1,0,0`: no move-ahead, (5) = 0 (the sample stage always compensates fully).
|
||||
|
||||
Changing the FZP range now involves, in addition to the text above:
|
||||
|
||||
- **`Feedforward`** (static FZP feedforward): `ff_limit`, `out_min` / `out_max` to the same values as the FZP `Ub` /
|
||||
`Lb` (they limit drive + feedforward).
|
||||
- **`FreqGen`** `out_min` / `out_max` (channels 0 / 1) limit the FZP drive only while an excitation is running
|
||||
(measurements).
|
||||
- **`OnTargetTrigger`** `WaitForSlew` and **`TargetSelect_x` / `_y`** `use_final`: keep the defaults (1 / 0) with a
|
||||
small range. `use_final` 1 + `WaitForSlew` 0 ("exposure during the motion") only makes sense when the FZP range
|
||||
covers the scan step; steps larger than the range then still wait for the sample stage.
|
||||
- Move-ahead (0) must fit into the range: at ±1.5 V at most ≈ −0.05 µm (x) — so in operation it stays 0.
|
||||
- After a change: web HMI tab **Loop check** level 1 (expected settings) and, with the beamline ready, the scan test
|
||||
(level 4: positioning time, stability during the exposure per step size).
|
||||