docs(flomni): FZP how-tos as Markdown, updated for the PREEMPT_RT loop, move-ahead renamed to move-beyond
Markdown versions of local_contact_only-howto_calibrate_the_fzp_scanner.pdf and local_contact_only-howto_limit_the_fzp_scanner_range.pdf (original text kept, update sections for the present flOMNI loop). New section "Move-beyond" (FAST_FZP_PID_LOOP(0) < 0 of the sample PIDs): mechanism, sizing (about half the FZP stroke), Abbe correction on the FZP side, hold (5) now reads the FZP drive incl. feedforward (flomni repo, branch fzp-move-beyond, 6a15f68). The PDFs are kept. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_019aWcwG89wi7oVyrefukXX6
This commit is contained in:
2 files changed
+224
No files matched your search
+183
@@ -0,0 +1,183 @@
|
||||
# How to calibrate the FZP scanner of flOMNI
|
||||
|
||||
Source of `local_contact_only-howto_calibrate_the_fzp_scanner.pdf`. The original text is kept unchanged; the
|
||||
section "Update 2026-10-20" at the end lists what the present flOMNI loop (PREEMPT_RT, 100 µs, branch `rt_preemt`)
|
||||
needs in addition (corrected 2026-10-23: "move-ahead" → **move-beyond**, see the section at the end).
|
||||
|
||||
```text
|
||||
This is a description of how to calibrate the FZP scanner of flOMNI
|
||||
|
||||
Rmemeber to do all modifications for both directions x and y!
|
||||
|
||||
1. Align flOMNI without the FZP scanner. This implies disabling the
|
||||
FZP scanner completely. To do so, in OrchestraHMI edit the parameters
|
||||
of the FZP scanner as follows.
|
||||
(a) Module PID_i_FZP: Set Param FAST_FZP_PID_LOOP(0) to 2. Then the FZP
|
||||
will just remain
|
||||
statically centered to the capacitive sensor position in closed loop.
|
||||
(b) Module PID_i: Set Param FAST_FZP_PID_LOOP(0) to 0. If this is
|
||||
negative, the FZP scanner will move ahead in a point-by-point scan by
|
||||
that amount.
|
||||
(c) Module PID_i: Set Param FAST_FZP_PID_LOOP(5) to 0. If this is
|
||||
larger than zero the sample stage motion is stopped if the position is
|
||||
reached by the FZP scanner (which is the case when the scanner is not
|
||||
active). Zero means thus full compensation by the sample stage.
|
||||
|
||||
In this state the setup can be aligned, and the affine matrix for
|
||||
ptychography can be found. This refers then purely to the alignment of
|
||||
the interferometer and X-ray detector.
|
||||
|
||||
2. In the next step we perform a scan (with a small beam and small
|
||||
range) such that only the FZP scanner is used and the sample stage
|
||||
will not move.
|
||||
|
||||
Before we start ensure that the Correction Matrix for the capacitive
|
||||
sensors is cleared.
|
||||
|
||||
To do this change the parameters of the flOMNI control
|
||||
loop as follows:
|
||||
|
||||
(a) Module AnalogInput_6259: Set parameter MatrixCorr to 1 0 0 1
|
||||
(b) Module PID_j_fzp: Set Param FAST_FZP_PID_LOOP(0) to 1 to re-activate
|
||||
the FZP scanner.
|
||||
(c) Module PID_i: Set Param FAST_FZP_PID_LOOP(5) to 10, so up to 10
|
||||
volts output of the fast PID is accepted before starting a movement of
|
||||
the sample stage.
|
||||
(d) Module PID_i: Param FAST_FZP_PID_LOOP(4) 0
|
||||
(e) Module PID_i: Param FAST_FZP_PID_LOOP(6) 1
|
||||
with d and e it will now stabilize the sample stage position at zero
|
||||
(f) Module PID_i: Param FAST_FZP_PID_LOOP(0) remains 0 (no move ahead
|
||||
in a scan)
|
||||
|
||||
With these settings perform a scan. Use a field of view matching the
|
||||
range of the FZP scanner. Watch the sample stage position in Orchestra
|
||||
to ensure it does not change!
|
||||
A good range should be 2.8 microns in x and 1.8 microns in y.
|
||||
Reconstruct this scan with Michal's position refinement code and ask
|
||||
for an update of the FZP matrix. This can then be put into the
|
||||
AnalogInput_6259 module.
|
||||
I guess one has to have the matrix found for the interferometer in the
|
||||
reconsturction code, such that the new matrix of the FZP scanner will
|
||||
adjust the coordinate system of the FZP scanner to the system of the
|
||||
interferometer. Later these positions are used together and should
|
||||
match in direction.
|
||||
|
||||
3. Back to default operating values:
|
||||
(a) Module PID_i: Set Param FAST_FZP_PID_LOOP(0) to -0.5, for up to
|
||||
0.5 microns move ahead in a scan.
|
||||
All other parameters are already as supposed to be.
|
||||
Default values for the sample stage pid, fast FZP parameter are "-
|
||||
0.5,0,0,0,0.1,0.5,0"
|
||||
Default values for the fzp pid, fast FZP parameter are "1,0.0001,0.1,-
|
||||
4,0,0,0"
|
||||
```
|
||||
|
||||
Naming: "move ahead" in the text above (sample PIDs `FAST_FZP_PID_LOOP(0)` < 0) is called **move-beyond** from
|
||||
2026-10-23 on: the sample stage moves *beyond* the scan point towards the next one, the FZP compensates and so sits
|
||||
beyond its centre during the exposure, on the side that gives it more stroke for the next step. It is not a
|
||||
"pre-move" of the FZP to the next point (that is `TargetSelect` `use_final` 1, see below).
|
||||
|
||||
## Update 2026-10-20 (flOMNI loop on PREEMPT_RT, mpc2844)
|
||||
|
||||
The loop has new modules since the text above was written (flOMNI `FeedbackLoop/tuning_guide.md`,
|
||||
`feedback_report.md` 5.6, 6n, 6o, 6q). Parameters are changed in the web HMI (`http://mpc2844:8080`, tab
|
||||
Parameters) instead of OrchestraHMI. `i` = x or y, as above. Module names now: `PID_x`, `PID_y` (sample stage),
|
||||
`PID_x_fzp`, `PID_y_fzp` (FZP scanner), `AnalogInput_6289` (was `AnalogInput_6259`).
|
||||
|
||||
**FAST_FZP_PID_LOOP has 7 elements** (`PID.cpp`). Sample stage PIDs: (0) move-beyond (negative, µm, point-by-point
|
||||
scans only; section "Move-beyond" below), (4) FZP error threshold (µm) and (5) FZP drive limit (V) for holding the sample stage while the FZP
|
||||
takes the step (5 = 0: never held), (6) 1 = hold the sample stage at 0. FZP PIDs: (0) mode (1 operation, 2 FZP
|
||||
centred on its cap sensor = disabled, 3 not commissioned, 4 debug), (1)–(3) min / max / gain of the old velocity
|
||||
correction (−4 before 2026-10-19, now 0: replaced by the feedforward).
|
||||
|
||||
**In addition to step 1 (FZP disabled):**
|
||||
|
||||
- `Feedforward`: `enable` 0,0 — the static FZP feedforward adds its drive also in FZP mode 2.
|
||||
- `AbbeCorrection`: `enable` 0 (default until the signs are known).
|
||||
- `TargetSelect_x`, `TargetSelect_y`: `use_final` 0; `OnTargetTrigger`: `WaitForSlew` 1 (defaults).
|
||||
- `FreqGen`: `enable` 0 (excitation for measurements; default).
|
||||
|
||||
**In addition to step 2 (FZP-only calibration scan):**
|
||||
|
||||
- FZP range: `PID_x_fzp` / `PID_y_fzp` `Ub` / `Lb` are ±1.5 V in operation (the cap sensors are calibrated over
|
||||
about ±2 V only). For the field of view above (2.8 × 1.8 µm) widen them, e.g. to ±10 V (FZP ≈ 0.13 µm/V in x,
|
||||
≈ 0.05 µm/V in y). The feedforward stays off (`enable` 0,0) during this scan.
|
||||
- The detector trigger (`OnTargetTrigger`) checks the relative error sample vs FZP (tolerance / hold time from the
|
||||
ConstEmitter, BEC `interferometer_trigger_*`, default 10 nm / 3 ms) and waits `MinDelay_ms` (15 ms).
|
||||
|
||||
**Step 3, default operating values (replace the values in the text above):**
|
||||
|
||||
| Module | Parameter | Value |
|
||||
|---|---|---|
|
||||
| `PID_x`, `PID_y` | `FAST_FZP_PID_LOOP` | `0,0,0,0,0.1,0,0` (no move-beyond: −0.5 µm is larger than the ±1.5 V FZP stroke, x ≈ ±0.19 µm, y ≈ ±0.08 µm) |
|
||||
| `PID_x_fzp` | `FAST_FZP_PID_LOOP`, `Ub` / `Lb` | `1,0.0001,0.1,0,0,0,0`, ±1.5 V |
|
||||
| `PID_y_fzp` | `FAST_FZP_PID_LOOP`, `Ub` / `Lb` | `1,0.0005,0.02,0,0,0,0`, ±1.5 V |
|
||||
| `Feedforward` | `enable`; `ff_limit`, `out_min` / `out_max` | 1,1; ±1.5 V (as the FZP `Ub` / `Lb`) |
|
||||
| `OnTargetTrigger` | `WaitForSlew`, `MinDelay_ms` | 1, 15 |
|
||||
| `TargetSelect_x`, `TargetSelect_y` | `use_final` | 0 |
|
||||
| `AbbeCorrection` | `enable` | 0 (until the signs are known) |
|
||||
| `AnalogInput_6289` | `MatrixCorr` | the new matrix from the calibration |
|
||||
|
||||
Check afterwards: web HMI tab **Loop check**, level 1 (it expects these values and lists live values that differ
|
||||
from the `.xmp` files); save the new values as default (Parameters → Unsaved changes) or in the `.xmp`.
|
||||
|
||||
**After the calibration (per experiment): faster scans, exposure during the motion.** With the FZP calibrated
|
||||
over a wider range the FZP can take the small steps alone (flOMNI `feedback_report.md` 6q, tested at ±7 V):
|
||||
|
||||
- `PID_x_fzp` / `PID_y_fzp` `Ub` / `Lb` and the `Feedforward` limits to the calibrated range (e.g. ±7 V);
|
||||
- `TargetSelect_x` / `_y` `use_final` 1 (the FZP aims at the new point at once, the sample stage follows its
|
||||
S-curve) and `OnTargetTrigger` `WaitForSlew` 0 (trigger on the relative error only); optionally `MinDelay_ms` 0;
|
||||
- optionally move-beyond, sized to the FZP stroke (≈ half of it, see below): at ±7 V (x ≈ ±0.9 µm, y ≈ ±0.36 µm)
|
||||
`PID_x` (0) e.g. −0.5, `PID_y` (0) e.g. −0.2 (the values tested in `feedback_report.md` 6q).
|
||||
|
||||
Measured with 101 ms exposures: 0.2–1 µm steps trigger after 4.5–5.8 ms instead of 17–32 ms; the error during the
|
||||
exposure grows with the step (1 µm: ≈ 6 nm max instead of 2.5 nm), so the setting depends on the experiment.
|
||||
|
||||
## Move-beyond (`PID_x` / `PID_y` `FAST_FZP_PID_LOOP(0)` < 0), 2026-10-23
|
||||
|
||||
**What it does** (`PID.cpp`, unchanged since the original loop). Only in point-by-point scans (scan state 2). The
|
||||
TrajectoryGenerator gives the next scan point (outputs 5 / 6, PID input 9). The sample PID then aims not at the
|
||||
target T but at T + m · sign(next − T), m = min(|FAST_FZP_PID_LOOP(0)|, |next − T|): the sample stage moves *beyond*
|
||||
the point towards the next one, by at most the step. The FZP loop keeps its own target T (it does not see m), so the
|
||||
FZP compensates the offset: during the exposure it holds the relative position at T while sitting at +m off its
|
||||
centre. At the next point the FZP then has stroke R + m in the step direction instead of R (R = stroke from the centre),
|
||||
so larger steps are taken by the FZP alone. Steps smaller than |FAST_FZP_PID_LOOP(0)|: the sample stage is one point
|
||||
ahead, the FZP moves between 0 and +step only. The offset is not shaped (InputSlew acts on T only); the sample stage
|
||||
moves to it with its own (slow) loop, i.e. partly during the exposure, and the FZP compensates that motion.
|
||||
|
||||
**Is it still the same in the new loop?** Yes. The relative position is computed as before:
|
||||
FZP error = Cap − (Chan − target) (mode 1), `Position sample vs FZP` = Chan − Cap (PID output 3, RT_sampler). The
|
||||
sample PIDs still get the shaped target (InputSlew) + move-beyond and the raw interferometer. The FZP side now has
|
||||
two additions that act on the FZP side only: `TargetSelect` (`use_final` 0 = shaped target as before, 1 = final
|
||||
target) and `AbbeCorrection` (Chan corrected for the stage tilt, see below). The `Feedforward` uses the same FZP
|
||||
target and Chan, so it supplies the +m offset of the FZP directly (static term m / K).
|
||||
|
||||
**Sizing.** During the exposure the FZP has to hold m + Abbe correction + noise / settling inside its stroke:
|
||||
|m| ≤ R − reserve, as a rule ≈ R / 2. R per axis = `Ub` × K (K ≈ 0.13 µm/V x, ≈ 0.05 µm/V y):
|
||||
|
||||
| FZP `Ub`/`Lb` | R x / y | move-beyond x / y (≈ R/2) |
|
||||
|---|---|---|
|
||||
| ±1.5 V (production) | ±0.19 / ±0.08 µm | ≈ −0.1 / ≈ −0.04 µm (production: 0) |
|
||||
| ±7 V (after calibration) | ±0.9 / ±0.36 µm | −0.5 / −0.2 µm (tested, 6q) |
|
||||
|
||||
The original default −0.5 µm in both axes dates from the ±9 V limits of the old loop (x ≈ ±1.2 µm, about R/2; in y
|
||||
±9 V ≈ ±0.45 µm, i.e. about the full stroke). When the FZP saturates while holding the offset, the relative error grows and the trigger
|
||||
waits.
|
||||
|
||||
**Abbe correction.** `AbbeCorrection` adds a new term to the measured sample x / y on the FZP side only:
|
||||
x = Chan2_x + θz · offset_y_for_x, y = Chan1_y + θz · offset_x_for_y (θz = Chan3 / 13 mm). So yes: it is an error of
|
||||
the sample x / y position (stage tilt about z, not seen by the x / y interferometer beams) that is corrected by the
|
||||
FZP x / y, not by the sample stage (`PID_x` / `PID_y` keep the raw Chan, the rotZ piezo is too slow). The positions
|
||||
recorded for the reconstruction (`Position sample vs FZP`) contain it as well. It uses FZP stroke (limited to
|
||||
`correction_limit_um` 0.1 µm per axis; y at rest ≈ 2 nm, ≈ 30–37 nm during 20 µm x steps) — keep this in the reserve
|
||||
of the move-beyond, in y it is a large part of the ±0.08 µm stroke. Keep it off (`enable` 0) during the FZP
|
||||
calibration scan (step 2).
|
||||
|
||||
**Notes.**
|
||||
- Hold of the sample stage, (4) / (5): input 8 of `PID_x` / `PID_y` was the FZP PID output *before* the
|
||||
`Feedforward` (with the feedforward on only the remainder, not the real FZP drive). Since 2026-10-23 it is the
|
||||
`Feedforward` output (FZP drive incl. feedforward), so (5) again compares the real FZP drive. No change with the
|
||||
feedforward off or with (5) = 0 (production).
|
||||
- On the last point of a scan the next point output is 0, so the move-beyond points towards 0 (as in the old loop).
|
||||
- `TargetSelect` `use_final` 1 is a different thing: the FZP aims at the new point at once while the sample stage
|
||||
follows the S-curve. Both can be combined (6q, E0).
|
||||
+41
@@ -0,0 +1,41 @@
|
||||
# How to limit the FZP scanner range of flOMNI
|
||||
|
||||
Source of `local_contact_only-howto_limit_the_fzp_scanner_range.pdf`. The original text is kept unchanged; the
|
||||
section "Update 2026-10-20" lists what the present flOMNI loop (PREEMPT_RT, 100 µs, branch `rt_preemt`) needs in
|
||||
addition. "ahead movement" (element 0) is called **move-beyond** from 2026-10-23 on: the sample stage moves beyond the
|
||||
scan point towards the next one, the FZP compensates and sits off-centre, which gives it stroke for the next step.
|
||||
|
||||
```text
|
||||
to limit the fzp scanner ranges
|
||||
in PID_ FAST FZP PID LOOP
|
||||
0: reduce to small ahead movement range, such as -0.05
|
||||
4: reduce error threshold, eg 0.01
|
||||
5: inform PID that FZP range is small, such as 0.1
|
||||
|
||||
in PID_ _FZP
|
||||
limit the analog output LB and UB
|
||||
```
|
||||
|
||||
## Update 2026-10-20 (flOMNI loop on PREEMPT_RT, mpc2844)
|
||||
|
||||
The limited range is the present operating state: `PID_x_fzp` / `PID_y_fzp` `Ub` / `Lb` = ±1.5 V because the cap
|
||||
sensors are calibrated over about ±2 V only (flOMNI `feedback_report.md` 6n). FZP ≈ 0.13 µm/V in x, ≈ 0.05 µm/V in y,
|
||||
so ±1.5 V ≈ ±0.19 µm (x) / ±0.08 µm (y). In the sample stage PIDs (`PID_x`, `PID_y`) `FAST_FZP_PID_LOOP` is
|
||||
`0,0,0,0,0.1,0,0`: no move-beyond, (5) = 0 (the sample stage always compensates fully).
|
||||
|
||||
Changing the FZP range now involves, in addition to the text above:
|
||||
|
||||
- **`Feedforward`** (static FZP feedforward): `ff_limit`, `out_min` / `out_max` to the same values as the FZP `Ub` /
|
||||
`Lb` (they limit drive + feedforward).
|
||||
- **`FreqGen`** `out_min` / `out_max` (channels 0 / 1) limit the FZP drive only while an excitation is running
|
||||
(measurements).
|
||||
- **`OnTargetTrigger`** `WaitForSlew` and **`TargetSelect_x` / `_y`** `use_final`: keep the defaults (1 / 0) with a
|
||||
small range. `use_final` 1 + `WaitForSlew` 0 ("exposure during the motion") only makes sense when the FZP range
|
||||
covers the scan step; steps larger than the range then still wait for the sample stage.
|
||||
- Move-beyond (0) must fit into the stroke: the FZP holds it during the exposure, together with the Abbe correction
|
||||
(up to `correction_limit_um` 0.1 µm when enabled) and noise. Rule ≈ half the stroke: at ±1.5 V ≈ −0.1 µm (x),
|
||||
≈ −0.04 µm (y); at ±7 V −0.5 / −0.2 µm (tested). Production: 0. Details: calibration how-to, section "Move-beyond".
|
||||
- (5) compares the real FZP drive incl. the feedforward (input 8 of `PID_x` / `PID_y` = `Feedforward` output since
|
||||
2026-10-23; before it was the FZP PID output without the feedforward).
|
||||
- After a change: web HMI tab **Loop check** level 1 (expected settings) and, with the beamline ready, the scan test
|
||||
(level 4: positioning time, stability during the exposure per step size).
|
||||