Update Calibrations

2024-03-15 17:27:00 +01:00
parent 0cd673b5eb
commit 59c3fc6521

@ -1,5 +1,7 @@
##### Probe semi-angle ##### Probe semi-angle
##### PAAR -
##### Detector camera length ##### Detector camera length
Pixelated STEM detector is located at a different position compared to level of nominal camera length (dedicated STEM detectors in our case). Real camera length to the pixelated detector has to be measured for each available nominal camera length. Otherwise wrong detector cover angle is used in reconstructions. Pixelated STEM detector is located at a different position compared to level of nominal camera length (dedicated STEM detectors in our case). Real camera length to the pixelated detector has to be measured for each available nominal camera length. Otherwise wrong detector cover angle is used in reconstructions.