Beamline hardware
PIPS diode plugged into cable 'PIPS1' and attached to central stage above JUNGFRAU detector via magnet Check cables are correct on the back of the PRIME chamber
Diagnostic plate should be installed and moved to knife edges
If you are operating in the tender X-ray regime, you will need to close and pump down the Prime chamber at this point
Back in the control room
Open ATT120 (icon is just upstream of APU120)
- Attenuate X-ray transmission to 0.000001
Alvra Launcher --> Prime --> MPOD SARES10-MPD1
- Open tab MPOD 2:EHS 84 20p
- Turn on ch-0 and ch-1 on to 50 V
Laser panel --> Open Prime Keysight --> Boxcar tab
- Check signal is present on the PIPS diode
From Alvra Beamline Synoptic, open KB mirrors, vertical (KBV123) and horizontal (KBV124)
To load previous settings
Alvra launcher --> Beamline - Aramis Synoptic In the Alvra endstation bit of the synoptic click the grey box 'SaveRes'
There is always a time stamp, but in the comments there is usually a description of the energy and the approximate beamsize.
Click on the one you want, then click 'Restore All'. You have to confirm two popup windows to really restore the values. The current vs. stored parameters are displayed below and have a 'not equal to' icon if they're different to one another. These will disappear once settings are restored.
To measure beamsize
This process is the same for x and y. But obviously makes use of
- KBv (KBV123) when measuring the vertical beamsize - uses horizontal knife edge and moves Huber in Y
- KBh (KBH124) when measuring the horizontal beamsize - uses vertical knife edge and moves Huber in Y
PIPS diode
Open the stripchart showing the readout of the PIPS diode to manually get to a position on the knife edge that is 'central' (i.e. beam half covered by knife edge).
Alvra launcher --> Tools --> pshell stripchart
- Open --> Keysight_Ch1_Ch2.scd
- You should now see the stripchart of the PIPS diode reading
Setup of scan
Use Huber stages to move the knife edge between open space and fully blocking the beam (use PIPS stripchart as readout/feedback)
Move until you are halfway between the two extremes
Open slic
- Use the 'scan' tab
- Select the relevant Prime motor motion from the dropdown (X or Y)
- Select 'Relative to current position'
- Enter +/- some number of microns (pick something sensible relative to the beamsize that makes sure you get to the two extremes (fully blocked and fully transmitting to diode)).
- Stepsize of 1 um
- 250 shots
Analysis of scan
Open a jupyter notebook of alvra_beamline_scripts Use notebook 'KnifeEdge_scan.ipynb'
Adjustment of beamsize
In KBv and KBh panels, adjust Bender 1 and Bender 2 in increments of 0.05.
You can check manually whether you have made the beam significantly smaller or larger by manually scanning in 1 um steps with Prime X or Prime Y on the Huber stages over the knife edge and looking at the signal on the PIPS and how many 'clicks' it takes to go from blocked to transmitting.
Remeasure the new beamsize by following 'To measure beamsize' instructions above.
Repeat measuring and adjusting iteratively until you get to the desired beamsize.
To save new settings
Alvra launcher --> Beamline - Aramis Synoptic In the Alvra endstation bit of the synoptic click the grey box 'SaveRes'
In top left hand area, add a comment to describe the condition
- X-ray energy
- Beamsize X
- Beamsize Y
- XOTA = 10 mm (with front time tool in, was = 0 mm when this wasn't there)
- 4 mrad is the standard (or 6 mrad)
Add labels from the drop down menu
Click 'Save'