Bernina Monochromator Scan
beale_e edited this page 2026-11-16 09:56:17 +01:00

Beam aligned to Bernina

Beam must be going to Bernina

  • In the Aramis synoptic, you should see that OMH064 has 'ESB' above it
  • To make this happen, open the 'Switch' drop down and select 'ESB'

Bernina Monochromator

Bernina Monochromator should be in

  • In the Bernina panel of the Aramis synoptic, click the button 'Bernina' at the bottom
  • From Beamline Mode tab select 'Fix Offset'
  • In the 'Energy Mono (eV)' box enter the desired energy in eV
  • There should be a 'MOVING' flag as the motion is underway

Bernina Monochromator Scan

The webapp can be reached at: https://sf-photodiag-test.psi.ch/app

Spectral tab (top) --> Calibration tab (under synoptic) --> Aramis mono scan tab (ESB)

  • You can use the webapp to configure everything by clicking 'prepare beamline (review first)'
  • Apply beamline changes - this will put the mono in at the energy of the machine
  • You can see in the green panel on the bottom right which checks it is doing, all are done automatically except the last one which is just to search and lock the optics hutch (so shutter can open)
  • Then need to set 'start energy' and 'stop energy' (you can use an downstream screen and manually move the mono energy to find where the intensity is highest/visible so that you can choose a sensible range)
  • Scan in 5 eV steps
  • Keep default 10 shots per point
  • Click 'start scan'

Remove Bernina Monochromator

Bernina Monochromator can be taken out

  • In the Bernina panel of the Aramis synoptic, click the button 'Bernina' at the bottom
  • From Beamline Mode tab select 'Pink
  • There should be a 'MOVING' flag as the motion is underway