Files
x07ma/script/HystScanMult.py
2020-02-18 13:34:57 +01:00

139 lines
4.8 KiB
Python

#Parameters
"""
FIELD = "Hx"
RANGES = [(-0.1, 0.1, 0,2),]
ENERGIES = (707.90, 703.90)
ENERGY_CHANGE_SLEEP = 0.5
MODE = 'CIRC +'
OFFSET = -1.0
"""
iPS = False # false for old power supply, True for new (Oxford Instr) supply
if iPS == True:
caput('X07MA-ES1-IPS:RAMP-ENBL', 1) # enable non-overshoot mode
FIELD_PRECISION = 0.01
if FIELD == field_z: FIELD = "Hz"
field = field_z if FIELD == "Hz" else (field_x_ips if iPS else field_x )
field_done = field_z_done if FIELD == "Hz" else (None if iPS else field_x_done)
ramp_rate = ramp_rate_z if FIELD == "Hz" else (ramp_rate_x_ips if iPS else ramp_rate_x)
START_FIELD=RANGES[0][0] # ADDED BY CP ON 7.2.2020
#Pre-actions
#cawait('ACOAU-ACCU:OP-MODE', 'Light Available', type = 's')
wait_beam()
if MODE is not None:
pol_mode.write(MODE)
if OFFSET is not None:
pol_offset.write(OFFSET) #caput('X07MA-ID:ENERGY-OFFS', OFFSET)
caputq('X07MA-PC:CSCALER.INPB', '1')
if len(ENERGIES) ==2:
dif_series = plot([],"Dif", title="Dif")[0].getSeries(0)
else:
dif_series = None
set_preference(Preference.ENABLED_PLOTS, ['field', 'tey_norm', 'trans_norm'])
set_preference(Preference.PLOT_TYPES, {'tey_norm':1, 'trans_norm':1})
print "Set energy..."
setpoint2 = ENERGIES[0]
print setpoint2
energy.write(setpoint2)
#sleep( ENERGY_CHANGE_SLEEP ) # Settling time
scan = ManualScan(['field', 'Energy'], ['TEY', 'I0', 'trans', 'polarization', 'polAngle', 'temperature', 'RingCurrent', 'fieldAnalogX', 'tey_norm','trans_norm'], [0.0, ENERGIES[0]], [0.0, ENERGIES[-1]], [0, len(ENERGIES)-1])
scan.start()
write_logs()
#for (START_FIELD, END_FIELD, RAMP_RATE) in RANGES:
#ramp_rate.write(RAMP_RATE)
# if abs(field.readback.read() - START_FIELD) > FIELD_PRECISION:
# print "Setting start field: ", START_FIELD
# field.write(START_FIELD)
# time.sleep(3.0)
#wait_device(field_done, 1) #ramp_done.wait_for_value(1.0)
if abs(field.readback.read() - START_FIELD) > FIELD_PRECISION:
ramp_rate.write(2.0)
field.write(START_FIELD)
if iPS == True:
caput("X07MA-PC-MAG:X:DMD",float(START_FIELD))
if field_done is None:
#ramp_rate.write(2.0)
ramp_rate.waitInPosition(2.0,10000)
time.sleep(1.0)
caput("X07MA-ES1-IPS:ACTION", 1)
time.sleep(3.0)
wait_device_in_range(field, START_FIELD, FIELD_PRECISION)
print "Waiting field done..."
while caget("X07MA-ES1-IPS:OUTPUT_RBV", "i") != 0 :
time.sleep(0.1)
print "Done waiting"
else:
time.sleep(3.0)
wait_device(field_done, 1) #ramp_done.wait_for_value(1.0)
#open_valve(2.0)
wait_device(pol_done, "DONE")
ramp_rate.write(RAMP_RATE)
print "Set end field: ", END_FIELD
field.write(END_FIELD) #caputq('X07MA-PC-PS2:M:GO.A', END_FIELD)
index = 0
while(True):
for setpoint2 in ENERGIES:
#Energy.put(setpoint2, timeout=None) # TODO: Set appropriate timeout
energy.write(setpoint2)
sleep( ENERGY_CHANGE_SLEEP ) # Settling time
#TODO: change to Controlled variable?
readback2 = energy_readback.read() #EnergyReadback.get()
if abs(readback2 - setpoint2) > 0.1 : # TODO: Check accuracy
raise Exception('Energy could not be set to the value ' + str(setpoint2))
#Detector field readback
field_readback = field.readback.read() #fieldReadback.get()
detector1 = signal_tey.read() #TEY.get()
detector2 = signal_i0.read() #I0.get()
detector3 = signal_trans.read() #trans.get()
detector4 = float(pol_mode.getPositions().index(pol_mode.readback.read())) #polarization.get()
detector5 = pol_angle.read() #polAngle.get()
detector6 = temperature.readback.read() #temperature.get()
detector7 = current.read()
detector8 = signal_field_analog_x.read() # fieldAnalogX.get()
_tey_norm = detector1/detector2
_trans_norm = detector3/detector2
if dif_series is not None:
if setpoint2 == ENERGIES[0]:
first = _tey_norm
else:
dif = _tey_norm-first
dif_series.appendData(field_readback,dif)
scan.append ([field_readback, setpoint2], [field_readback, readback2], [detector1, detector2, detector3, detector4, detector5, detector6, detector7, detector8, _tey_norm, _trans_norm])
after_sample() #To call check_id_error()
if field_done.read() == 1: #If not ramping #ramp_done.get() == 1:
print "Reached end field. Readback=", field_readback
break
index = index+1
scan.end()
if RAMP_RATE is not None:
ramp_rate.write(2.0)
ramp_rate.waitInPosition(2.0,10000)