Files
dev/script/HystScan.py
2018-01-19 10:56:53 +01:00

158 lines
5.3 KiB
Python
Executable File

#Script imported from: Fe_hyst_plus.xml
#Parameters
START_FIELD = -0.2
END_FIELD = 0.2
ENERGIES = (707.90, 703.90)
RAMP_RATE = 2.0
ENERGY_CHANGE_SLEEP = 0.5
#Pre-actions
#cawait('ACOAU-ACCU:OP-MODE', 'Light Available', type = 's')
#caput('X07MA-ID:MODE', 'CIRC +')
sleep(1.0)
#cawait('X07MA-ID:DONE', 'DONE', ype = 's')
#caput('X07MA-ID:ENERGY-OFFS', '-1.0')
#cawait('X07MA-ID:DONE', 'DONE', type = 's')
caputq('X07MA-PC:CSCALER.INPB', '1')
caputq('X07MA-PC-PS2:SET:DMD:RAMPRATE:TPM', RAMP_RATE)
#sleep(15.0)
if len(ENERGIES) ==2:
dif_series = plot([],"Dif", title="Dif")[0].getSeries(0)
else:
dif_series = None
set_preference(Preference.ENABLED_PLOTS, ['field', 'tey_norm', 'trans_norm'])
set_preference(Preference.PLOT_TYPES, {'tey_norm':1, 'trans_norm':1})
#scan = ManualScan(['field', 'Energy'], ['TEY', 'I0', 'trans', 'polarization', 'polAngle', 'temperature', 'RingCurrent', 'fieldAnalogX', 'tey_norm','trans_norm'])
scan = ManualScan(['field', 'Energy'], ['TEY', 'I0', 'trans', 'polarization', 'polAngle', 'temperature', 'RingCurrent', 'fieldAnalogX', 'tey_norm','trans_norm'], [0.0, ENERGIES[0]], [0.0, ENERGIES[-1]], [0, len(ENERGIES)-1])
scan.start()
#Stop condition
ramp_done = Channel('X07MA-PC-MAG:X:RAMP:DONE', type = 'i')
#Creating channels: dimension 1
#RegionPositioner field
#field = Channel('X07MA-PC:GO', type = 'd')
fieldReadback = Channel('X07MA-PC-PS2:STS:PMCURRENT', type = 'd')
#Creating channels: dimension 2
#ArrayPositioner Energy
Energy = sim_energy #Channel('X07MA-PHS-E:GO.A', type = 'd')
EnergyReadback = sim_energy_readback # Channel('X07MA-PGM:CERBK', type = 'd')
#ScalarDetector TEY
TEY = Channel('X07MA-ES1-AI:SIGNAL0', type = 'd')
#ScalarDetector I0
I0 = Channel('X07MA-ES1-AI:SIGNAL1', type = 'd')
#ScalarDetector trans
trans = Channel('X07MA-ES1-AI:SIGNAL2', type = 'd')
#ScalarDetector polarization
#polarization = Channel('X07MA-ID:MODE', type = 'd')
#ScalarDetector polAngle
#polAngle = Channel('X07MA-ID:ALPHA', type = 'd')
#ScalarDetector temperature
temperature = Channel('X07MA-PC-TC:STS:T1', type = 'd')
#ScalarDetector RingCurrent
RingCurrent = Channel('ARIDI-PCT:CURRENT', type = 'd')
#ScalarDetector fieldAnalogX
fieldAnalogX = Channel('X07MA-ES1-AI:SIGNAL4', type = 'd')
print "Waiting for start field"
"""
#WORKAROUND TO BUG IN DRIVER
if abs(fieldReadback.get() - START_FIELD) < 0.005:
caput('X07MA-PC-PS2:M:GO.A', START_FIELD + 0.01)
caput('X07MA-PC-PS2:M:GO.A', START_FIELD)
#caputq('X07MA-PC-PS2:M:GO.A', START_FIELD)
"""
caput("X07MA-PC-MAG:X:DMD", START_FIELD)
caput("X07MA-PC-MAG:STARTRAMP.PROC", 1)
ramp_done.wait_for_value(1.0)
print "Set end field"
caput("X07MA-PC-MAG:X:DMD", END_FIELD)
caputq("X07MA-PC-MAG:STARTRAMP.PROC", 1)
index = 0
while(True):
#ArrayPositioner Energy
for setpoint2 in ENERGIES:
Energy.put(setpoint2, timeout=None) # TODO: Set appropriate timeout
sleep( ENERGY_CHANGE_SLEEP ) # Settling time
readback2 = EnergyReadback.get()
if abs(readback2 - setpoint2) > 0.1 : # TODO: Check accuracy
raise Exception('Actor Energy could not be set to the value ' + str(setpoint2))
#Detector field readback
field_readback = fieldReadback.get()
#Detector TEY
detector1 = TEY.get()
#Detector I0
detector2 = I0.get()
#Detector trans
detector3 = trans.get()
#Detector polarization
detector4 = sim_energy_readback.read() # polarization.get()
#Detector polAngle
detector5 = sim_energy_readback.read() #polAngle.get()
#Detector temperature
detector6 = temperature.get()
#Detector RingCurrent
detector7 = RingCurrent.get()
#Detector fieldAnalogX
detector8 = fieldAnalogX.get()
#Manipulation tey_norm
#Variable Mappings
#TODO: Move, if needed, this import to the file header: import math
tey_norm = detector1/detector2
#Manipulation trans_norm
#Variable Mappings
#TODO: Move, if needed, this import to the file header: import math
trans_norm = detector3/detector2
if dif_series is not None:
if setpoint2 == ENERGIES[0]:
first = tey_norm
else:
dif = tey_norm-first
dif_series.appendData(field_readback,dif)
print [field_readback, setpoint2]
print [field_readback, readback2]
print [detector1, detector2, detector3, detector4, detector5, detector6, detector7, detector8, tey_norm, trans_norm]
scan.append ([field_readback, setpoint2], [field_readback, readback2], [detector1, detector2, detector3, detector4, detector5, detector6, detector7, detector8, tey_norm, trans_norm])
print "Field = " + str(fieldReadback.get())
if ramp_done.get() == 1: #If not ramping
print "Not ramping, breaking execution"
break
index = index+1
#Closing channels
Energy.close()
EnergyReadback.close()
TEY.close()
I0.close()
trans.close()
#polarization.close()
#polAngle.close()
temperature.close()
RingCurrent.close()
fieldAnalogX.close()
#field.close()
fieldReadback.close()
ramp_done.close()
scan.end()