# Manual to Make the Calibration Update ## Changing the Range from 10 µA to 100 µA This guide describes the required hardware and software changes to update the calibration range from **10 µA** to **100 µA**. --- # 1. Hardware Update Replace the resistor values on the PCB from **2 kΩ** to **200 Ω**. ### New Resistor Part Number - **RNCF0603TKY200R** ### Resistors to Replace The following resistors must be changed: - R10 - R28 - R46 - R64 - R82 - R100 - R118 - R136 --- # 2. Software Update After completing the hardware modification, update the device configuration using the provided Python script. ## Install Requirements Run the following command to install all required Python packages: ```bash pip install -r .\requirements.txt ``` ## Upload the New Scale Values Use the script: ```bash MTB_UploadNewScale.py ``` ### Procedure 1. Enter the device serial number. 2. Depending on your country, enter the correct meanCount value: - 240 for Europe - 250 for the USA 3. The script uploads the new configuration automatically. ## Calibration Update Complete After the upload finishes successfully, the device is configured for the new 100 µA range.