diff --git a/ReadMe.md b/ReadMe.md index 263d07a..36585d2 100644 --- a/ReadMe.md +++ b/ReadMe.md @@ -1,13 +1,62 @@ -# Manual to make the Calibration Update (changing the range from 10uA to 100uA) +# Manual to Make the Calibration Update -For installing the requirements use: +## Changing the Range from 10 µA to 100 µA -``` +This guide describes the required hardware and software changes to update the calibration range from **10 µA** to **100 µA**. + +--- + +# 1. Hardware Update + +Replace the resistor values on the PCB from **2 kΩ** to **200 Ω**. + +### New Resistor Part Number + +- **RNCF0603TKY200R** + +### Resistors to Replace + +The following resistors must be changed: + +- R10 +- R28 +- R46 +- R64 +- R82 +- R100 +- R118 +- R136 + +--- + +# 2. Software Update + +After completing the hardware modification, update the device configuration using the provided Python script. + +## Install Requirements + +Run the following command to install all required Python packages: + +```bash pip install -r .\requirements.txt ``` -Use the `MTB_UploadNewScale.py` script to upload the new scale values. +## Upload the New Scale Values -1. Enter your serial number -2. depending on you country, enter the meanCount (240 for europe and 250 for the usa) -3. It uploads the configuration +Use the script: + +```bash +MTB_UploadNewScale.py +``` + +### Procedure + +1. Enter the device serial number. +2. Depending on your country, enter the correct meanCount value: + - 240 for Europe + - 250 for the USA +3. The script uploads the new configuration automatically. + +## Calibration Update Complete + +After the upload finishes successfully, the device is configured for the new 100 µA range.