This page contains technical details of PSSS device. For user guide see [PSSS user guide](Aramis spectral diagnostics/PSSS user guide). # Camera settings **25.08.20** - Camera sever: 73 - Event: 14 - Pulser: 1 - Total delay: 6500.994 (offset -1) +1000 = -2499.006 us when not at 100 Hz, or 7500.994 at 100 Hz - Width: 5.000 us - Camera: PCO.edge 5.5 16 bit - Board number: 0 - Exposure: 9 ms, 0 delay - ROI: 2560 x 2160 (x,y) **31.1.21** Changes from above so camera is always triggered at 100 Hz - Exposure: 9.9 ms, 0 delay - Event: 38 - Pulser: 5 - Total delay: 0.0 + 6650.000 us - Width: 5.000 us **16.6.23** **Pulse ID buffering is no longer working!** Correct pulse ID confirmed at 50 Hz - Camera event: 14 - Pulser: 6 - Total delay: 1035.937 + 6500.994 us - Width: 7.000 us - BS event:75 - ROI: 2560 x 1162 (500:1661) x,y # Drawings - [Camera and screen assembly](http://planarchiv.psi.ch/plandb_pub/p_plandb_cus_pub_sucw.viewItem?pId=211061&pPosInQueryList=1&pCount=7840&pIndexInQueryList=728&pKrz=&pHg=50024&pUg=&pLnr=&pAnlage=&pBaugruppe=&pBenennung=&pFremdNr=&pGueltig=GUL) - [Main spectrometer overview](http://planarchiv.psi.ch/plandb_pub/p_plandb_cus_pub_sucw.viewItem?pId=209837&pPosInQueryList=1&pCount=1&pIndexInQueryList=1&pKrz=&pHg=50024&pUg=21&pLnr=420&pAnlage=&pBaugruppe=&pBenennung=&pFremdNr=&pGueltig=GUL) - [Grating arm](http://planarchiv.psi.ch/plandb_pub/p_plandb_cus_pub_sucw.viewItem?pId=209245&pPosInQueryList=1&pCount=1090&pIndexInQueryList=291&pKrz=&pHg=50024&pUg=21&pLnr=&pAnlage=&pBaugruppe=&pBenennung=&pFremdNr=&pGueltig=GUL) - [Grating chamber](http://planarchiv.psi.ch/plandb_pub/p_plandb_cus_pub_sucw.viewItem?pId=208882&pPosInQueryList=1&pCount=1090&pIndexInQueryList=283&pKrz=&pHg=50024&pUg=21&pLnr=&pAnlage=&pBaugruppe=&pBenennung=&pFremdNr=&pGueltig=GUL) - [Calibration filter holder](http://planarchiv.psi.ch/plandb_pub/p_plandb_cus_pub_sucw.viewItem?pId=210203&pPosInQueryList=1&pCount=1090&pIndexInQueryList=337&pKrz=&pHg=50024&pUg=21&pLnr=&pAnlage=&pBaugruppe=&pBenennung=&pFremdNr=&pGueltig=GUL) # Motor positions Motor set points for grating and crystal positions. Crystal type -SARFE10-PSSS059:MOTOR_X3 | Crystal| Readback | | ------ | ------ | | None| - 0.0| | Si(111)R155| -29.0 | | Si(220)R75| -51.0| | Si(220)R145| -73.0| | Si(220)R200| -82.6 (low limit switch)| Grid type - SARFE10-PSSS055:MOTOR_X1 | Grating| Readback | | ------ | ------ | | None| -0.0| | 100 nm a| -62.0 | | 100 nm b| -86.0| | 150 nm | -110.0| | 200 nm | -134.0| # Look-up tables **Girder position** - moves the PSSS into the position of the diffracted order **Si(111) Radius of curvature 155** - rotates crystal, sets camera position and rotation and sets camera calibration **Si(220) Radius of curvature 75** - rotates crystal, sets camera position and rotation and sets camera calibration **Si(220) Radius of curvature 145** - rotates crystal, sets camera position and rotation and sets camera calibration **Si(220) Radius of curvature 200** - rotates crystal, sets camera position and rotation and sets camera calibration