Update SATOP31 PSRD175

This commit is contained in:
arrell_c 2023-11-03 11:52:49 +01:00
parent db4cd58736
commit a874e24508

View File

@ -7,7 +7,7 @@ Double stack, [F2224-21P159](https://www.hamamatsu.com/eu/en/product/optical-sen
# Imaging
- Camera:[ACA1300-60GM CS](https://www.baslerweb.com/de-de/shop/aca1300-60gc/?creative=532311824379&keyword=%2Baca1300-60gc&matchtype=b&network=g&device=c&gad_source=1&gclid=Cj0KCQjwtJKqBhCaARIsAN_yS_mp7pHmpzlgwy-e5SafMKNrnMCxLcLLwKBQtCkyAo_QDMZDA9PuK8kaAn_gEALw_wcB)
- Lens: 35 mm [MVL35M23](https://www.thorlabs.com/thorproduct.cfm?partnumber=MVL35M23)
- Camera serial number: 22470860
- Camera serial number: 21574220
# High voltage set points
- MCP_front set point -1500 V
- Screen set point 3000 V
@ -15,28 +15,18 @@ Double stack, [F2224-21P159](https://www.hamamatsu.com/eu/en/product/optical-sen
Voltage ramp time from turn on/off approx 2 minutes
**Turn on**
_MCP front_: `caput SAROP21-PSRD103:CH0-ONOFF-SP 1`
_Screen_: `caput SAROP21-PSRD103:CH1-ONOFF-SP 1`
_MCP front_: `caput SATOP31-PSRD175:CH0-ONOFF-SP 1`
_Screen_: `caput SATOP31-PSRD175:CH1-ONOFF-SP 1`
**Turn off**
_MCP front_: `caput SAROP21-PSRD103:CH0-ONOFF-SP 0`
_Screen_: `caput SAROP21-PSRD103:CH1-ONOFF-SP 0`
_MCP front_: `caput SATOP31-PSRD175:CH0-ONOFF-SP 0`
_Screen_: `caput SATOP31-PSRD175:CH1-ONOFF-SP 0`
# Config
- Image preview PV : SAROP21-PSRD103:FPICTURE
- Trigger currently from SGE-CPCW-74-EVR0, FrontUnivOut9
# Camera view
The camera does not image the full MCP screen (aprrox 70 %). Below are screen shots explaining what is seen by the camera.
<img src="https://docs.google.com/drawings/d/e/2PACX-1vQgQt1JOrCNkeBEhYpXwS-DoMsCCCMa3sHWzQP350plvV5mK2mYEx_KYJrRPkmZRw3Lo4GhOrZh1eHv/pub?w=1402&amp;h=1054">
<img src="https://docs.google.com/drawings/d/e/2PACX-1vRVXM3PQYnclY8MU40jpGMQCGvim2UZSzZNMe4JA3MBj2OYsuv4xARf07iDXrPUr92jMOA0L2ux5EL1/pub?w=1412&amp;h=1067">
<img src="https://docs.google.com/drawings/d/e/2PACX-1vTJtlLUAfUyu3onqnlmqR18ZcpYhhZPs_xWpcbjv1hCoZ3CpApn9AKMjpj7kBfSU0chnkOwlMvNRPCq/pub?w=1391&amp;h=1038">
- Image preview PV : SATOP31-PSRD175:FPICTURE
- Trigger currently from SGE-CPCW-B1-EVR0, FrontUnivOut11
# Triggering
**29.06.20**
- Event: 13
- Pulser: 0
- Total delay: 6498.999 + 500 = 6998.999 us
**03.11.23**
- Event: 52
- Pulser: 6
- Total delay: 6499.006 + 1003.011 = 7502.017 us
- Width: 5.000 us
<img src="https://docs.google.com/drawings/d/e/2PACX-1vQ-9_DmsnDC6xGJEK0-whVVjRSSjTe8AEv1BNXuPdxvBMqMBoQJ4E3g3EGTsw6Vr4AzgjEi2LwH-51Z/pub?w=1391&amp;h=684">