Update PSSS user guide

This commit is contained in:
arrell_c 2023-08-23 13:51:29 +02:00
parent d4b9d63fa3
commit 9ef4dbf923

View File

@ -30,6 +30,8 @@ The most common reason for the PSSS spectrum not to be updating is the processin
Occasionally the background image ID is not listed. If this is the case a new background image should be taken. To do this first set the camera ROI to `2560x1162` (required for 100 Hz operation :disappointed:), then press the capture button in the panel shown above. The generated image ID then needs to be copied and updated in the config as described above. Occasionally the background image ID is not listed. If this is the case a new background image should be taken. To do this first set the camera ROI to `2560x1162` (required for 100 Hz operation :disappointed:), then press the capture button in the panel shown above. The generated image ID then needs to be copied and updated in the config as described above.
As the camera chip is read from the center out in the y direction the ROI settings are symmetric around the center. To set an ROI or 2560x1162 first enter 500 in the from y pix box and press Set Parameter. The ROI readback will then show you the applied ROI (hopefully 2560x1162).
Still have an issue? Use the contact list for the PSSS here:[contacts](Aramis Contacts) Still have an issue? Use the contact list for the PSSS here:[contacts](Aramis Contacts)
## Photon energy outside window ## Photon energy outside window