Update PBPS calibration guide

This commit is contained in:
arrell_c 2025-02-13 13:44:02 +01:00
parent cfeca59bde
commit 395981f7f1

View File

@ -1,27 +1,104 @@
## Operating principle
Single shot diagnostic to measure relative pulse energy and beam position. Back-scattered X-rays from a thin target (typically 50 nm silicon nitride) are measured by four photodiodes positioned above, below, left and right of the X-ray beam.
## Operating Principle
The PBPS (Photon Beam Position Solid) is a **single-shot diagnostic** used to measure **relative pulse energy and beam position**. It functions by detecting **back-scattered X-rays** from a thin target, typically a **50 nm silicon nitride (Si3N4) film**. The scattered X-rays are captured by four **photodiodes** arranged around the beam:
Position data maintains FEL pointing, while single-shot pulse energy measurements are used to normalise experiment data. There is one PBPS in the front-end and two further devices per end station.
- **Top** (Up)
- **Bottom** (Down)
- **Left**
- **Right**
## How to use
These measurements are critical for:
- **FEL pointing stability** (by tracking beam position)
- **Experimental data normalization** (by measuring pulse energy)
SwissFEL features **one PBPS in the front-end** and **two PBPS devices per end station**.
---
## How to Use
### **1. Open the Photon Diagnostics Web App**
[Click here to open](https://sf-photodiag.psi.ch/app)
### **2. Select the PBPS Device**
- Navigate to the **Position** or **Calibration** tab.
- In the **first dropdown menu**, select the PBPS device you want to calibrate.
### **3. Choose the Target**
- In the **second dropdown menu**, select a target.
- **Default target:** **50nm Si3N4**.
### **4. Start Calibration**
- Press the **Calibrate** button.
- The system will:
- Normalize the response of the **four photodiodes**.
- **Scan the PBPS device** horizontally and vertically across the FEL beam.
- **Update the plots** with the calibration response (takes ~1 minute).
### **5. Analyze the Results**
- The updated plots show:
- **X-axis** → Device motor position (**Horizontal: left plot**, **Vertical: right plot**)
- **Y-axis** → Normalized **horizontal or vertical diode pair response**
- **A good calibration** results in a **straight line** passing through the three scan points.
### **6. Push the Calibration Results**
- If the calibration looks good, press **"Push results/elog"**.
- This updates the data processing pipeline and generates an **elog entry**.
- After an **automated restart (~30s)**, the **PV and BS position & intensity data** will be updated with the new calibration.
1. Open photon diagnostics [web app](https://sf-photodiag.psi.ch/app)
2. Under Position/calibration tabs,s elect device from drop down menu 1
3. Select target from drop down menu 2 - default target is 50nm Si3N4
4. Press calibrate
<img src="https://docs.google.com/drawings/d/e/2PACX-1vRGynTQ54ugugA_bM9pYc9eZDlet3Z9MSZWn17wJNwAEM2ORiJGoL8O61frpLBVRpFKlzJ7M05a5Mb1/pub?w=1245&h=1039">
---
The calibration procedure normalizes the response of the 4 diodes, then scans the device horizontally and vertically over the FEL beam to calibrate the position response.
Once completed (approx 1 minute) the plots are updated with the calibration response. Plotted on the x axis is the device motor position (horizontal - left plot and vertical - right plot) and on the y axis is plotted the normalized horizontal or vertical diode pair response. For a good setup the plot should be a straight line through the 3 data points.
If you are happy with the calibration, press the "Push results/elog" button where the new calibration data is sent to the required data processing pipelines. After an automated restart (~30 seconds) both PV and BS position and intensity data will be calculated with the new calibration. An elog entry is made detailing the calibration data.
## Error
**Soft limits**
Error message: `ERROR: Motor position outside soft limits`
Problem is a soft limit on either the device Motor X or Y is blocking the scan motion
1. Select all from drop down menu for motor X and motor Y of PBPS GUI (shown below)
2. Change motor soft limits to 1 and -1 for both motors
3. Retry calibration
## Troubleshooting
### **Soft Limits Error**
#### **Error Message:**
`ERROR: Motor position outside soft limits`
#### **Cause:**
- The PBPS **Motor X or Y** has hit a **soft limit**, preventing the scan.
#### **Solution:**
1. **Open the PBPS GUI** and locate the motor settings.
2. **Modify the soft limits**:
- Set **Motor X soft limits** to `-1` and `1`
- Set **Motor Y soft limits** to `-1` and `1`
3. **Retry Calibration**.
<img src="https://docs.google.com/drawings/d/e/2PACX-1vSga-Zluo4b0isNs92C5MXfawMueNVk3ideRRDA_ZGQmD05Jn4noTS_pOl0OHiOyjS31OhLTAelFkYn/pub?w=728&amp;h=686">
### **Additional Debugging Tips**
- Ensure the **PBPS device is correctly selected**.
- Verify that the **target is installed properly**.
- Try manually moving the motor **in small steps** to check functionality.
- If the issue persists, check for **elog messages** indicating motor faults.
---
## FAQ
**Q: How do I know if my calibration was successful?**
A: If the diode response fits a straight line and the y-axis range is a few percent, your calibration is good!
**Q: What happens if I push bad calibration results?**
A: The **old calibration** will be overwritten, which may cause **incorrect beam position data**. Make sure to verify results before pushing.
**Q: Why does calibration fail even after adjusting soft limits?**
A: The device may be in a fault state. Try **resetting the motor controller** or checking the motor motion is completing by checking the motor GUI.
---
## 🎭 **Do Say / Don't Say**
**Do Say:**
- *"I just calibrated PBPS like a ninja!"*
- *"Lets push these results to elog!"*
**Don't Say:**
- *"Whats a soft limit?"*
- *"Do I really need to calibrate this thing?"*
---