## Operating principle
|
||||||
|
Single shot diagnostic to measure relative pulse energy and beam position. Back-scattered X-rays from a thin target (typically 50 nm silicon nitride) are measured by four photodiodes positioned above, below, left and right of the X-ray beam.
|
||||||
|
|
||||||
|
Position data maintains FEL pointing, while single-shot pulse energy measurements are used to normalise experiment data. There is one PBPS in the front-end and two further devices per end station.
|
||||||
|
|
||||||
|
## How to use
|
||||||
|
|
||||||
|
1. Open photon diagnostics [web app](https://sf-photodiag.psi.ch/app)
|
||||||
|