From 03eb40a4ccf06b0f5ef03e391d0ebe46bdf74e24 Mon Sep 17 00:00:00 2001 From: arrell_c Date: Fri, 16 Jun 2023 17:21:29 +0200 Subject: [PATCH] Update SARFE10 PSSS059 --- Aramis-spectral-diagnostics/SARFE10-PSSS059.md | 10 ++++++++++ 1 file changed, 10 insertions(+) diff --git a/Aramis-spectral-diagnostics/SARFE10-PSSS059.md b/Aramis-spectral-diagnostics/SARFE10-PSSS059.md index fa1b509..096ca18 100644 --- a/Aramis-spectral-diagnostics/SARFE10-PSSS059.md +++ b/Aramis-spectral-diagnostics/SARFE10-PSSS059.md @@ -21,6 +21,16 @@ Changes from above so camera is always triggered at 100 Hz - Total delay: 0.0 + 6650.000 us - Width: 5.000 us +**16.6.23** +**Pulse ID buffering is no longer working!** +Correct pulse ID confirmed at 50 Hz +- Camera event: 14 +- Pulser: 6 +- Total delay: 1035.937 + 6500.994 us +- Width: 7.000 us +- BS event:75 +- ROI: 2560 x 1162 (500:1661) x,y + # Drawings - [Camera and screen assembly](http://planarchiv.psi.ch/plandb_pub/p_plandb_cus_pub_sucw.viewItem?pId=211061&pPosInQueryList=1&pCount=7840&pIndexInQueryList=728&pKrz=&pHg=50024&pUg=&pLnr=&pAnlage=&pBaugruppe=&pBenennung=&pFremdNr=&pGueltig=GUL) - [Main spectrometer overview](http://planarchiv.psi.ch/plandb_pub/p_plandb_cus_pub_sucw.viewItem?pId=209837&pPosInQueryList=1&pCount=1&pIndexInQueryList=1&pKrz=&pHg=50024&pUg=21&pLnr=420&pAnlage=&pBaugruppe=&pBenennung=&pFremdNr=&pGueltig=GUL)