// SPDX-FileCopyrightText: 2026 Filip Leonarski, Paul Scherrer Institute // SPDX-License-Identifier: GPL-3.0-only #ifndef JUNGFRAUJOCH_SENSORABSORPTION_H #define JUNGFRAUJOCH_SENSORABSORPTION_H #include #include // How much of the beam a flat sensor stops, as a function of the angle at which the beam enters it. // // A photon arriving at incidence angle alpha to the sensor normal crosses t/cos(alpha) of sensor // instead of t, so more of it is absorbed: the detector is MORE efficient at high angle than at // normal incidence, and an uncorrected reflection out at the detector edge reads high. Dividing by // that ratio is the whole correction. // // a(alpha) = t / (L cos alpha) crossing length in attenuation lengths // QE(alpha) = 1 - exp(-a) absorbed fraction // correction to I = QE(0) / QE(alpha) normalised so normal incidence is untouched // // Two limits, both physical and both reached in this corpus: // - L << t (long wavelength): every photon stops in the entrance skin, QE = 1 at every angle and // the correction is exactly 1. This is why the low-energy case is inert rather than gated off. // - L >> t (thin sensor, hard X-rays): absorption is proportional to path length and the // correction tends to cos(alpha) - a factor of 2 at 2theta = 60 degrees. This is the regime // small-molecule work at 18-25 keV lives in. // // alpha is the angle to the DETECTOR NORMAL, not the scattering angle. The two coincide only on an // untilted detector. On a tilted one the correction acquires an azimuthal dependence at fixed // resolution, which is the only part of it that is not degenerate with an overall Wilson B. // // Attenuation coefficients: NIST X-Ray Mass Attenuation Coefficients (Hubbell & Seltzer), total // mu/rho with coherent scattering. Photoelectric absorption dominates over the whole range used // here, and because the correction is a RATIO of two absorbed fractions the photoelectric branching // ratio cancels exactly; only the attenuation length enters. Coherent scattering, which attenuates // without converting, is a few per cent of mu/rho in Si below 20 keV and is kept in the total - it // biases L slightly short, and the correction depends on L only through t/L. // // NOT modelled, and negligible where this is used: K-fluorescence escape from the sensor (in CdTe // above the Cd K edge at 26.71 keV a fluorescence photon can leave the sensor, so the charge is // recorded in the wrong pixel or not at all); charge sharing between pixels; and the obliquity of // the entrance window. Above the Cd/Te K edges the escape term is large and this model should not // be trusted as it stands. namespace sensor_absorption { // NIST mass attenuation coefficient mu/rho [cm^2/g] against photon energy [keV]. Absorption edges // are the repeated energies - the table is walked in order and interpolated log-log, which is how // these tables are meant to be read. Si has only the K edge at 1.839 keV, far below anything used. struct MuRow { double E_keV, mu_rho; }; inline constexpr MuRow SI_TABLE[] = { {1.0, 1570.0}, {1.5, 535.5}, {1.8389, 309.2}, {1.8389, 3192.0}, {2.0, 2777.0}, {3.0, 978.4}, {4.0, 452.9}, {5.0, 245.0}, {6.0, 147.0}, {8.0, 64.68}, {10.0, 33.89}, {15.0, 10.34}, {20.0, 4.464}, {30.0, 1.436}, {40.0, 0.7012}, {50.0, 0.4385}}; inline constexpr MuRow CD_TABLE[] = { {1.0, 7350.0}, {1.5, 2931.0}, {2.0, 1473.0}, {3.0, 541.4}, {3.5375, 357.5}, {3.5375, 1152.0}, {3.63101, 1083.0}, {3.727, 1013.0}, {3.727, 1389.0}, {4.0, 1170.0}, {4.018, 1157.0}, {4.018, 1324.0}, {5.0, 768.5}, {6.0, 479.3}, {8.0, 225.4}, {10.0, 124.4}, {15.0, 41.78}, {20.0, 19.20}, {26.7112, 8.809}, {26.7112, 50.65}, {30.0, 37.65}, {40.0, 17.78}, {50.0, 9.779}}; inline constexpr MuRow TE_TABLE[] = { {1.0, 8434.0}, {1.5, 3608.0}, {2.0, 1832.0}, {3.0, 679.2}, {4.0, 329.7}, {4.3414, 267.8}, {4.3414, 788.2}, {4.47465, 750.4}, {4.612, 699.5}, {4.612, 944.5}, {4.7728, 878.2}, {4.9392, 806.2}, {4.9392, 929.2}, {5.0, 901.4}, {6.0, 572.1}, {8.0, 270.2}, {10.0, 150.1}, {15.0, 50.78}, {20.0, 23.41}, {30.0, 7.878}, {31.8138, 6.738}, {31.8138, 37.19}, {40.0, 20.64}, {50.0, 11.45}}; // Log-log interpolation, clamped at both ends of the table. template double InterpMuRho(const MuRow (&tab)[N], double E_keV) { if (E_keV <= tab[0].E_keV) return tab[0].mu_rho; if (E_keV >= tab[N - 1].E_keV) return tab[N - 1].mu_rho; std::size_t i = 1; while (i < N - 1 && tab[i].E_keV < E_keV) i++; const double e0 = tab[i - 1].E_keV, e1 = tab[i].E_keV; if (!(e1 > e0)) // the two rows of an absorption edge: take the upper side return tab[i].mu_rho; const double f = (std::log(E_keV) - std::log(e0)) / (std::log(e1) - std::log(e0)); return std::exp(std::log(tab[i - 1].mu_rho) + f * (std::log(tab[i].mu_rho) - std::log(tab[i - 1].mu_rho))); } // Attenuation length 1/mu [um] of a sensor material at a given wavelength. Si and CdTe are the // sensors in use; an unrecognised material is treated as silicon, which is what DetectorSetup // defaults to anyway. inline double AttenuationLength_um(const std::string &material, double lambda_A) { if (!(lambda_A > 0.0)) return 0.0; const double E_keV = 12.39842 / lambda_A; double mu_rho_cm2_g, rho_g_cm3; if (material == "CdTe") { // Mass fractions from the atomic weights, Cd 112.414 and Te 127.60. constexpr double w_cd = 112.414 / 240.014, w_te = 127.60 / 240.014; mu_rho_cm2_g = w_cd * InterpMuRho(CD_TABLE, E_keV) + w_te * InterpMuRho(TE_TABLE, E_keV); rho_g_cm3 = 5.85; } else { mu_rho_cm2_g = InterpMuRho(SI_TABLE, E_keV); rho_g_cm3 = 2.3290; } const double mu_cm = mu_rho_cm2_g * rho_g_cm3; return mu_cm > 0.0 ? 1e4 / mu_cm : 0.0; } // Everything the per-reflection correction needs, reduced to the two numbers that are constant for // a dataset. Built once on the host; the GPU predictor takes the two floats. struct SensorQE { float a0 = 0.0f; // t / L, the optical thickness at normal incidence float qe0 = 1.0f; // absorbed fraction at normal incidence bool active = false; // false where the sensor is opaque and the correction is exactly 1 // Opaque beyond this: exp(-20) = 2e-9, so QE(0)/QE(alpha) rounds to exactly 1.0f for every // incidence angle and the correction is bit-identical to not applying it. This is what makes // the long-wavelength case inert without a flag or a threshold anyone has to choose. static constexpr float OPAQUE_A0 = 20.0f; static SensorQE Build(const std::string &material, double thickness_um, double lambda_A) { SensorQE q; const double L = AttenuationLength_um(material, lambda_A); if (!(thickness_um > 0.0) || !(L > 0.0)) return q; q.a0 = static_cast(thickness_um / L); q.qe0 = static_cast(1.0 - std::exp(-q.a0)); q.active = q.a0 < OPAQUE_A0; return q; } // The multiplicative correction to an intensity recorded at incidence angle alpha: QE(0)/QE(alpha). // Always <= 1, because a sensor is more efficient off-normal than head-on. [[nodiscard]] float Factor(float cos_alpha) const { if (!active || !(cos_alpha > 1e-3f)) return 1.0f; const float qe = 1.0f - std::exp(-a0 / cos_alpha); return qe > 0.0f ? qe0 / qe : 1.0f; } }; } // namespace sensor_absorption #endif // JUNGFRAUJOCH_SENSORABSORPTION_H