api: record the beam size at the sample, and write it where NXmx puts it

dataset_settings gains beam_size_x_um and beam_size_y_um, the horizontal and
vertical size of the X-ray beam where it meets the sample. They follow the same
route total_flux takes - OpenAPI, DatasetSettings, the CBOR start message, the
HDF5 master, and back out of a stored file - and nothing consumes them; this is
metadata a beamline can state and a downstream program can read.

NXmx puts this in the application definition rather than the base class: not
NXbeam's extent (rank 2, nP x 2, per scan point, always FWHM of a rectangular
aperture) but NXmx's own incident_beam_size, a recommended rank-1 two-element
array in the order x, y. Both are live and neither is deprecated, so the choice
matters; the MX definition wins in an MX file. Written as one array with a
units attribute of "m", like every other length in the master, so the settings
hold micrometres and FillMessage converts once.

The unit table of ReadLength_m becomes LengthUnitFactor so the array read can
share it: a master written elsewhere may state this in millimetres.

Co-Authored-By: Claude Opus 5 (1M context) <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01EFEJG6WBQv8th4UJFNe53N
This commit is contained in:
leonarski_fandClaude Opus 5 committed 2026-09-02 09:59:34 +02:00
1 parent 7890149ad6
commit 8a04773d1d
20 files changed
+253 -10

No files matched your search

+2
View File
@@ -33,6 +33,8 @@ There are minor differences at the moment:
| incident_energy | float | X-ray energy \[eV\] | X |
| incident_wavelength | float | X-ray wavelength \[Angstrom\] | X |
| incident_wavelength_spread | float (optional) | FWHM of the X-ray wavelength distribution \[Angstrom\] (NXmx incident_wavelength_spread); omitted when the beam is monochromatic | |
| beam_size_x | float (optional) | Horizontal size of the X-ray beam at the sample \[m\] (first element of NXmx incident_beam_size) | |
| beam_size_y | float (optional) | Vertical size of the X-ray beam at the sample \[m\] (second element of NXmx incident_beam_size) | |
| frame_time | float | Frame time, if multiple frames per trigger \[s\] | X |
| count_time | float | Exposure time \[s\] | X |
| saturation_value | int64 | Maximum valid sample value | X |
+1
View File
@@ -24,6 +24,7 @@
* `jfjoch_writer` writes `direct_beam_x`/`direct_beam_y` in the HDF5 master - where the undeflected beam lands - beside the `beam_center_x`/`beam_center_y` PONI.
* `jfjoch_writer` writes `/entry/MX/peakCountUnfiltered` in the HDF5 master beside the other per-image spot counts, instead of only in the data files.
* `jfjoch_broker` sends `direct_beam_x`/`direct_beam_y` on the CBOR start message.
* `dataset_settings` takes `beam_size_x_um`/`beam_size_y_um`, the size of the X-ray beam at the sample, and `jfjoch_writer` writes them as `incident_beam_size` in the HDF5 master.
* `jfjoch_viewer` reads PILATUS miniCBF sweeps natively, and draws grid scan cells in the proportion of the scan steps.
* The rugnux manual is reorganised into task pages with a run overview, and gains worked phenix / REFMAC5 / POINTLESS-AIMLESS / careless examples.
* `jfjoch_viewer` labels the merge-statistics plot over the range the axis is drawn on, so the CC1/2 curve is no longer read against tick labels covering only part of it.
+1
View File
@@ -154,6 +154,7 @@ File-level HDF5 attributes `file_name`, `file_time`, `HDF5_Version` are also set
| `incident_wavelength` | NXmx | angstrom |
| `incident_wavelength_spread` | NXmx | angstrom (only if polychromatic) |
| `total_flux` | NXmx | Hz |
| `incident_beam_size` | NXmx | m (two elements, x then y; written only when both beam sizes are given) |
### `/entry/instrument/attenuator` (NXattenuator)