<?xml version="1.0" encoding="UTF-8"?>
|
|
<?oxygen RNGSchema="http://www.oasis-open.org/docbook/xml/5.0/rng/docbook.rng" type="xml"?>
|
|
<chapter xmlns="http://docbook.org/ns/docbook" xmlns:xlink="http://www.w3.org/1999/xlink"
|
|
version="5.0">
|
|
<info><title>Rheometer</title><author>
|
|
<personname>Ferdi Franceschini</personname>
|
|
</author>
|
|
<date/>
|
|
</info>
|
|
<sect1>
|
|
<title>Configuration</title>
|
|
<para>The driver is loaded into SICS by adding the following line in the
|
|
<literal>/usr/local/sics/extraconfig.tcl</literal> file </para>
|
|
<para>
|
|
<variablelist>
|
|
<varlistentry>
|
|
<term><command>add_rheo name IP tol settletime PORT</command></term>
|
|
<listitem>
|
|
<para><replaceable>name</replaceable> of parameter from rheometer, use
|
|
either <command>rhSpeed</command> or <command>rhTorque</command> for
|
|
consistency with existing data files</para>
|
|
<para><replaceable>IP</replaceable> is the address of Moxa box</para>
|
|
<para><replaceable>tol</replaceable> is the tolerance of
|
|
<replaceable>name</replaceable></para>
|
|
<para><replaceable>settletime</replaceable> is the time to wait in seconds
|
|
before checking that we have reached the set value of
|
|
<replaceable>name</replaceable>.</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
</variablelist>
|
|
</para>
|
|
<para><command>add_rheo rhSpeed ca3-quokka 0.01 5 4001</command></para>
|
|
<para>This will create a driver with a tolerance of 0.01 for the speed, with a settle time
|
|
of 5 seconds. </para>
|
|
<para>The rheometer speed and torque will be available under the <literal>/sample</literal>
|
|
group and will be saved in the data file as <literal>../sample/rhTorque</literal> and
|
|
<literal>../sample/rhSpeed</literal>.</para>
|
|
<para>A standard HISTOGRAM_XY file (ie a QKKxxxx.nx.hdf XY data file)will be created by
|
|
default. Use <command>newfile</command> to specify other filetypes. </para>
|
|
</sect1>
|
|
<sect1>
|
|
<title>Commands</title>
|
|
<para>The rheometer driver in SICS starts an acquisition on the histogram memory when the
|
|
rheometer matches the speeds in a list of trigger values held by the driver. Data will
|
|
be saved after each acquisition. </para>
|
|
<para>
|
|
<variablelist>
|
|
<varlistentry>
|
|
<term><command>hset /sample/rhSpeed/runexp 1</command></term>
|
|
<listitem>
|
|
<para>The driver begins a sequence when this flag is set to 1. When the flag
|
|
is 0 (default) the rheometer speed is ignored and no sequence is run.
|
|
The flag is automatically reset to 0 at the end of a sequence</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
</variablelist>
|
|
</para>
|
|
</sect1>
|
|
<sect1>
|
|
<title>Parameters</title>
|
|
<para>
|
|
<variablelist>
|
|
<varlistentry>
|
|
<term><command>hset /sample/rhTorque</command></term>
|
|
<listitem>
|
|
<para>
|
|
<option>scale </option>(default 1) scale the voltage reading to the
|
|
torque value</para>
|
|
<para><option>offset </option>(default 0) applies an offset if
|
|
necessary.</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>hset /sample/rhSpeed</command></term>
|
|
<listitem>
|
|
<para><option>scale</option> (default 1) scale the voltage reading to the
|
|
speed value</para>
|
|
<para><option>offset</option> (default 0) applies an offset if necessary.</para>
|
|
<para><option>saveIndex</option> The index of the next entry to be saved in
|
|
the data file, starts at zero</para>
|
|
<para><option>triggerList</option> (default "EMPTY") List of speeds which
|
|
will trigger an acquisition and save data</para>
|
|
<para><option>acqTime</option> (default "EMPTY") List of histogram memory
|
|
acquisition times. If you only put one value here it will be applied to
|
|
all of the acquisitions</para>
|
|
<para><option>runexp</option> (default 0) The driver ignores the rheometer
|
|
speed until this flag has been set to 1. NOTE: It is automatically reset
|
|
at the end of a sequence</para>
|
|
<para><option>tol</option> Tolerance, an acquisition will be triggered when
|
|
the rheometer speed is within tolerance of the current trigger value in
|
|
the triggerList. </para>
|
|
<para>The <option>triggerList</option> and <option>acqTime</option> lists
|
|
will be cleared when a sequence has been completed.</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
</variablelist>
|
|
</para>
|
|
<example>
|
|
<title>Rheometer example</title>
|
|
<para>
|
|
<literallayout>
|
|
<computeroutput>
|
|
hset /sample/rhSpeed/triggerList 4 5 6 5 4
|
|
OK
|
|
hset /sample/rhSpeed/acqTime 5 10 15 10 5
|
|
OK
|
|
hset /sample/rhSpeed/runexp 1
|
|
OK
|
|
Data will be saved in a standard HISTOGRAM_XY file
|
|
You can override this default by running the 'newfile' command first
|
|
|
|
rhCallBack /sics/rhSpeed 3.9807 trigger a 5 second acquisition,
|
|
sicstime 2011-06-10 17:33:35
|
|
rheometer_savehmmdata /sics/rhSpeed: save data index = 0,
|
|
sicstime 2011-06-10 17:33:35
|
|
rhCallBack /sics/rhSpeed 4.9562 trigger a 10 second acquisition,
|
|
sicstime 2011-06-10 17:33:43
|
|
rheometer_savehmmdata /sics/rhSpeed: save data index = 1,
|
|
sicstime 2011-06-10 17:33:43
|
|
rhCallBack /sics/rhSpeed 5.98 trigger a 15 second acquisition,
|
|
sicstime 2011-06-10 17:33:50
|
|
rheometer_savehmmdata /sics/rhSpeed: save data index = 2,
|
|
sicstime 2011-06-10 17:33:50
|
|
rhCallBack /sics/rhSpeed 5.023 trigger a 10 second acquisition,
|
|
sicstime 2011-06-10 17:34:06
|
|
rheometer_savehmmdata /sics/rhSpeed: save data index = 3,
|
|
sicstime 2011-06-10 17:34:06
|
|
rhCallBack /sics/rhSpeed 4.0226 trigger a 5 second acquisition,
|
|
sicstime 2011-06-10 17:34:20
|
|
rheometer_savehmmdata /sics/rhSpeed: save data index = 4,
|
|
sicstime 2011-06-10 17:34:20
|
|
</computeroutput>
|
|