Files
sics/site_ansto/instrument/sans/config/anticollider/acscript.txt
2013-02-20 18:19:44 +11:00

6 lines
265 B
Plaintext

# Forbid detector motion when the detector voltage is on
forbid {-inf inf} for det when dhv1 in {800 inf}
forbid {-inf inf} for detoff when dhv1 in {800 inf}
forbid {-inf 625} for det when hez in {-inf 647}
forbid {-inf inf} for hez when det in {-inf 625}